![YOKOGAWA ADMAG TI Series Скачать руководство пользователя страница 64](http://html.mh-extra.com/html/yokogawa/admag-ti-series/admag-ti-series_installation-manual_911189064.webp)
<7. Errors and Countermeasures (Display unit)>
63
IM 01E24A01-01EN
7. Errors and Countermeasures
(Display unit)
The error messages are described in the following tables.
For AXFA11, read the applicable user’s manual as listed in Table 1.1.
Explanation of NE107 status:
NE107 status
Status of the device
F
Failure
Device malfunction, Parts malfunction
C
Function Check
The output signal is temporarily invalid for the local operation or manual operation.
S
Out of Specification
The device works in out of specification.
The output signal is uncertain for the process or the ambience.
M
Maintenance Required The maintenance is required in the near future.
N
No Effect
Other issue
System Alarm
Device breaks down and causes abnormal measurement.
Device replacement is needed.
NE107
Status
Error Message
Error Description
Countermeasure
Display
F
010:Main CPU FAIL
CPU (Main board) failure was detected.
Contact Yokogawa service center.
F
011:Rev calc FAIL
Failure of reverse calculation was detected.
Contact Yokogawa service center.
F
012:Main EEP FAIL
Failure of EEPROM (Main board) was detected.
Turn on the power again within the temperature range. If the problem
does not improve, contact Yokogawa service center.
F
013:Main EEP dflt
EEPROM (Main board) was reseted to default values.
Contact Yokogawa service center.
F
014:Snsr bd FAIL
Failure of sensor board was detected.
Contact Yokogawa service center.
F
015:Snsr comm ERR
Communication error of sensor was detected.
Contact Yokogawa service center.
F
016:AD 1 FAIL[Sig]
Failure of A/D transumitter 1 [flow velocity signal]
wasdetected.
Contact Yokogawa service center.
F
017:AD 2 FAIL[Excit]
Failure of A/D transmitter 2 [Exciting current] was detected. Contact Yokogawa service center.
F
018:Coil open
Coil of sensor was disconnected.
Turn off the power, check coil of sensor and excitation cable.
F
019:Coil short
Coil of sensor was shorted.
Contact Yokogawa service center.
F
020:Exciter FAIL
Failure of excitation circuit was detected.
Contact Yokogawa service center.
F
021:PWM 1 stop
Error of pulse width modulation 1 was detected.
Contact Yokogawa service center.
F
022:PWM 2 stop
Error of pulse width modulation 2 was detected.
Contact Yokogawa service center.
F
023:Opt bd mismatch
Mismatch of option board was detected.
Contact Yokogawa service center.
F
024:Opt bd EEP FAIL
Failure of EEPROM (option board) was detected.
Contact Yokogawa service center.
F
025:Opt bd A/D FAIL
Failure of A/D (option board) was detected.
Contact Yokogawa service center.
F
026:Opt bd SPI FAIL
Failure of SPI (option board) was detected.
Contact Yokogawa service center.
F
027:Restore FAIL
Restore of parameters was failed.
Retry parameter restoration.
F
028:Ind bd FAIL
Failure of indicator board was detected.
Check the ambient temperature of display is within the range. If the
problem does not improve, contact Yokogawa service center.
F
029:Ind bd EEP FAIL
Failure of EEPROM (indicator board) was detected.
Turn on the power again within the temperature range. If the problem
does not improve, contact Yokogawa service center.
F
030:LCD drv FAIL
Failure of LCD driver was detected.
Contact Yokogawa service center.
F
031:Ind bd mismatch
Mismatch of Indicator board was detected.
Contact Yokogawa service center.
F
032:Ind comm ERR
Communication error of indicator board was detected.
Check connection of Indicator & main board.
F
033:microSD FAIL
Failure of microSD card was detected.
Change microSD card.
F
244:Amp EEP FAIL
Amplifier EEPROM failed.
Replace electrical parts such as the amplifier. Or replace the device.
C
249:SoftDL Incmplt
Software download is incomplete.
Check the cables, power supply and MTB SoftDL Error (MTB.
SOFTDL_ERROR)
C
250:SoftDL FAIL
Software download has failed.
Check the download file and MTB SoftDL Error (MTB.SOFTDL_
ERROR)
F
251:Abnml Boot PRS
Abnormal boot processing was detected at the startingtime.
Check the cables, power and MTB SoftDL Error (MTB.SOFTDL_
ERROR)
F
335:IT1 ttl bkup ERR
IT1 Total Backup has failed. Last IT1 Output.Value (IT1.
LAST.VALUE) is not saved.
Contact the nearest sales office or service center.
F
339:IT2 ttl bkup ERR
IT2 Total Backup has failed. Last IT2 Output.Value (IT2.
LAST.VALUE) is not saved.
Contact the nearest sales office or service center.
Errors and Countermeasures (Display unit)
7