<5. Functions>
144
IM 01E21A02-03EN
5.11 Event Management Function
5.11.1 Storing Parameter (Backup)
The backup function can store parameter settings in the display’s built-in memory (display
board) or microSD card (with the optional code MC selected). If the backup data is restored, the
parameter settings can be duplicated to another product. The display’s built-in memory can store
data for three backups and the microSD card can store as much data as the capacity allows.
There are three backup methods available: Backup from the main board of this product to the
memory on the display board, backup from the main board of the product to the microSD card,
and backup from the memory on the display board to the microSD card. The file name, backup
name, and date can be specified using the backup function.
The data backed up to the microSD card is stored in the “YOKOGAWA” folder as a “.PAR” file.
The setting can be configured with the following parameters.
For details about the backup parameters, refer to Subsection 5.11.3.
Menu path
Display
Device setup ► Diag/Service ► Param bkup/restore ► (see below)
F
OUNDATION
Fieldbus
Device Configuration ► MTB ► Device Configuration ► Maintenance ► Parameter
Backup/Restore ► (see below)
Block
Name
Relative
Index
Parameter
Description
Display
F
OUNDATION
Fieldbus
MTB
72
F backup name Factory Backup Name
Displays the backup name defined
at shipment from the manufacturing
factory.
MTB
73
F backup date
Factory Backup Date
Displays the backup date at
shipment from the manufacturing
factory.
MTB
74
SD backup name SD Backup Name
Specifies the name of the file to be
backed up to the microSD card.
MTB
75
Backup name 1 Backup Name 1
Specifies backup name 1.
Up to 16 characters
MTB
76
Backup date 1
Backup Date 1
Specifies date 1.
MTB
77
Backup name 2 Backup Name 2
Specifies backup name 2.
Up to 16 characters
MTB
78
Backup date 2
Backup Date 2
Specifies date 2.
MTB
79
Backup name 3 Backup Name 3
Specifies backup name 3.
Up to 16 characters
MTB
80
Backup date 3
Backup Date 3
Specifies date 3.
MTB
81
Backup execute Backup Execute
Specifies the execution of the
backup function.
*1
MTB
82
Backup result
Backup Result
Displays the result of the backup
function.
*2