![YOKOGAWA 7556 Скачать руководство пользователя страница 39](http://html.mh-extra.com/html/yokogawa/7556/7556_user-manual_911164039.webp)
4-8
IM 755601-01E
4.5 Setting the Measurement Time
Keys
NORMAL
MEAS&
LOCK
SAMPLE
EXTERNAL
REFERENCE
MANUAL
FAST
HI
HI LIMIT
EXIT
SETUP/MEAS
CHECK
MISC
I/F
MEAS
TIME
STORE
DATA
RECALL
DATA
LOCAL
ENTER
LIMIT
MODE
TRIG
MODE
SHIFT
R/%
REF
LIMIT
TRIG
7
1
0
.
2
3
5
4
6
8
9
LO LIMIT
IN
LO
M
Ω
M
Ω
k
Ω
Ω
,%
k
Ω
Ω
Ω
%
%
HIGH
SPEED
STORE
DATA
RECALL
DATA
CHECK
REMOTE
The appropriate indicator turns ON
Procedure
The items that are specified or selected are confirmed when the ENTER (
Ω
,%) key is
pressed. When confirmed, the display returns to the top menu of the SETUP mode.
To exit from a menu in the middle of the operation, press the EXIT (R/%) key. The
display returns to the top menu of the SETUP mode.
Switching to the SETUP mode
1.
Press the
SHIFT
SETUP/MEAS
R/%
key to switch to the SETUP mode. If the “MEAS & LOCK”
indicator is off, this operation is not necessary.
Selecting the measurement time
2.
3.
SHIFT
REF
LIMIT
ENTER
Ω
,%
[Main display]
MEAS
TIME
7
Explanation
Measurement time
Select from the following choices.
Type
Measurement Time* (when the measurement range is 1
Ω
to 1 M
Ω
)
nor (NORMAL)
19.9 ms (for 60 Hz)
23.2 ms (for 50 Hz)
FASt (FAST)
5.7 ms
H-SP (HIGH SPEED) 2.8 ms
* Measurement time: When the trigger mode is set to EXTERNAL, the time from the trigger
input to the falling edge of the EOM signal of the handler interface is
called the measurement time.
When the trigger mode is set to Manual or Internal, the EOM signal is not
output.
When the contact check function is ON (before the measurement), add 2 ms.
When the contact check function is ON (after the measurement), add 1 ms.
When a trigger delay is specified, add the delay time.
When the measurement range is 10 M
Ω
, add 4 ms.
When the measurement range is 100 M
Ω
, add 50 ms.