12.1 Input Specifications
For 755611 (1 digit = 0.001%)
• Accuracy 90 days after adjustment
Range
NORMAL
FAST
HIGH SPEED
1
Ω
0.012+10
0.012+20
0.012+30
10
Ω
0.01+3
0.01+10
0.01+20
100
Ω
0.008+3
0.008+10
0.008+20
1 k
Ω
0.006+3
0.006+10
0.006+20
10 k
Ω
0.006+3
0.006+10
0.006+20
100 k
Ω
0.006+3
0.006+10
0.006+20
1 M
Ω
0.01+4
0.01+20
0.01+40
10 M
Ω
0.03+10
0.3+20
0.3+40
100 M
Ω
0.2+20
• Accuracy one year after adjustment
Range
NORMAL
FAST
HIGH SPEED
1
Ω
0.015+10
0.015+20
0.015+30
10
Ω
0.012+3
0.012+10
0.012+20
100
Ω
0.011+3
0.011+10
0.011+20
1 k
Ω
0.009+3
0.009+10
0.009+20
10 k
Ω
0.009+3
0.009+10
0.009+20
100 k
Ω
0.009+3
0.009+10
0.009+20
1 M
Ω
0.015+4
0.015+20
0.015+40
10 M
Ω
0.04+10
0.3+20
0.3+40
100 M
Ω
0.2+20
Accuracy at 1
Ω
range measuring 0.1
Ω
or less (% of r
Ω
) is as follows.
90 days after adjustment
NORMAL:
0.012+0.0001
FAST:
0.012+0.0002
HIGH SPEED: 0.012+0.0003
One year after adjustment
NORMAL:
0.015+0.0001
FAST:
0.015+0.0002
HIGH SPEED: : 0.015+0.0003
Accuracy under the following conditions.
•
±
9.99% mode (10 times the digit error of the accuracy shown above for
±
99.9% mode)
• 23
±
5
°
C
• Add a temperature coefficient at 5 to 18 and 28 to 40
°
C range.
• Temperature coefficient :
±
[1/10 the measurement accuracy (the accuracy one year after
adjustment)]/
°
C
• Only the NORMAL mode is specified for the 100 M
Ω
range
• Excludes influence from the lead wires
Conditions to satisfy the specifications
• Lead wire resistance of H
CUR
and L
CUR
×
test current
≤
1.5 V
• Lead wire resistance of H
POT
and L
POT
≤
1.5
Ω
(1
Ω
range)
≤
30
Ω
(other than 1
Ω
range)
Maximum Common Mode Voltage
Between each measurement input terminal and case: 42 Vpeak