Connecting to Ground
When measuring signals that contain high frequency components, to
reduce the ringing caused by the ground lead inductance, it is important
to reduce the length of the ground connection. Affix the included Cu
sheet after adjusting its size to match the size of the device under
measurement. Then connect it to the ground (ground potential) by
soldering or by other means. Good ground connection can be made using
the included ground blade or ground spring.
Ground blade etc.
Cu sheet
To Ground pin
Device
Probe main unit
IC cap
Installing the Pincher Tip
To use the pincher tip, first install it on the probe so that the pincher tip
comes to the line indicated by the arrow when pulled.
Pincher tip
Probe main unit
Arrow and line indicated
on the probe body
3. How to Use