Operating Procedures
Preparation
1.
Have the probe and a digital oscilloscope with a probe interface ready.
2.
Insert the probe interface completely into the oscilloscope input, and
confirm that the BNC connector and interface pin are securely fastened.
You will hear the latch click when the connectors lock into place.
3.
When you connect the probe to a digital oscilloscope with a probe
interface, the probe’s attenuation ratio and input coupling are set
automatically.
4.
Attach any of the provided attachments or attachments that you
constructed to the probe head signal input terminals.
Note
If the digital oscilloscope does not detect the probe automatically, manually configure
the following settings on the oscilloscope.
• Set the probe attenuation ratio to 50:1.
• Set the input coupling to 50 Ω.