15-5
IM 701730-01E
Other Operations
15
15.3 Measuring the Offset Voltage and Applying the
Offset Voltage to the Computed Results
Procedure
ACTION
DELAY
X - Y
MENU
MENU
PHASE
SEARCH
HORIZONTAL
VERTICAL
TRIGGER
TRIG D
HELP
ESC
SELECT
RESET
SIMPLE
MODE
POSITION
MATH
MISC
FILE
CLEAR
TRACE
HISTORY
MEASURE
SETUP
PRESET
DISPLAY
CURSOR
CH
1
CH
2
CH
3
CH
4
POWER
COPY IMAGE SAVE
SHIFT
ZOOM
GO/NO-GO
ENHANCED
ACQ
START/STOP
SNAP
SHOT
V DIV
TIME DIV
• To exit the menu during operation, press
ESC
located above the soft keys.
• For a description of the operation using a USB
keyboard or a USB mouse, see section 4.3.
1.
Press the
MISC
.
2.
Press the
System Cnfg
soft key to display a configuration menu.
3.
Press the
Offset Cancel
soft key to select ON or OFF.
The default setting is OFF.
Explanation
The offset value that is specified for each channel can be subtracted from the input
signal and used for computation and automated measurement.
OFF: The offset value is not applied to computed results and the results of automated
measurements.
Waveform is observed without subtracting the offset voltage (DC voltage) from
the input signal.
The vertical position of the screen corresponds to the offset voltage.
ON:
The offset value is applied to computed results and the results of automated
measurements.
The offset value specified for each channel can be used to subtract the offset
voltage (DC voltage) from the input signal for waveform observation. The vertical
position is set to 0 V.
Note
The Offset Cancel setting applies to all channels.
To specify whether to cancel or not cancel the offset voltage for each channel separately, use
the linear scaling function.