13-19
IM 701730-01E
Ethernet Communications (Optional)
13
Displaying the Log List
8.
Press the
Connect Log List
soft key. The data and time, user name, and IP
address of the twenty-four recent accesses are displayed.
Note
•
The instrument supports two clients, but file operations cannot be performed simultaneously.
•
When the instrument is accessed from the PC or workstation (login),
is displayed at the
upper right corner of the screen.
•
This function cannot be used when using the FTP client function, LPR client function, or Web
server function or when performing file operations.
•
The log list is cleared when the power is turned OFF.
•
To use this function, you must configure TCP/IP in advance according to the procedure given
in section 13.2.
•
To activate the settings, you must power cycle the instrument.
Explanation
The floppy disk drive or PC card on the instrument can be accessed from a PC or
workstation on the network via the Ethernet network.
To access these drives, FTP client software is needed on the PC or workstation.
User Account
• User Name
You can enter the password using up to fifteen characters. The default setting is
“anonymous.” The characters that can be used are all the ASCII characters on the
keyboard.
• Password
You can enter the password using up to fifteen characters. The characters that can be
used are all the ASCII characters on the keyboard.
• Time Out
The connection to the network is automatically closed if there is no access to the
instrument for the specified time. The default setting is 5 s. Set the timeout to a long
value when using dedicated FTP client software.
13.6 Accessing the Instrument from a PC or Workstation (FTP Server Function)