10-7
IM 701730-01E
Analyzing and Searching Waveforms
10
Selecting the Search Logic
19.
Press the
Logic
soft key to select AND or OR.
Executing/Aborting the Search
20.
Press the
Search Exec
soft key. The search is executed, and the words Search
Exec change to Search Abort.
To abort the search, press the
Search Abort
soft key. The search is aborted, and
the words Search Abort change to Search Exec.
Displaying the Found Waveform
21.
Press
ESC
. The screen returns to the HISTORY menu.
22.
Display the waveform according to the procedure given in section 10.1.
The history map lists the record numbers and time stamps of the searched
waveforms.
Resetting the Search Results
23.
Set the Search Mode to OFF in the HISTORY menu, or turn OFF Zone1 to Zone4
and execute the search. The search result is reset.
Explanation
This section explains the setup procedures for searching history waveforms that pass or
do not pass a specified search zone.
Search Zone: Select Zone
You can register up to four search zones, Zone1 to Zone4. For each search zone, you
can set the search conditions below. You can also set whether to use AND or OR logic
of the search conditions of the four search parameters to perform the search (see
“Search Logic” described later).
• Status of the search parameter for the searched waveform
• Waveform to search
• Search zone range
Status of the Search Parameter for the Searched Waveform: Condition
You can select the condition of the target waveform to be searched in the search zone.
OFF: No search is performed.
IN:
Searches waveforms that pass through the specified search zone. Also searches
cases when the waveforms pass through the boundary lines.
OUT: Searches waveforms that do not pass through the specified search zone.
Waveform to Search: Source
You can select the search target waveform. If a waveform that meets the search
condition is found, other history waveforms at the same time as the found waveform are
also displayed.
• For the DL1735E/DL1740E/DL1740EL
You can select the target waveform from CH1 to CH4, Math1, and Math2.
• For the DL1720E
You can select from CH1 to CH2, and Math1.
10.2 Searching the History Waveforms Using Zones (History Search)