7-2
IM 701730-01E
7.2
Setting the Record Length
<For a description of this function, refer to page 2-14.>
Procedure
ACTION
DELAY
X - Y
MENU
MENU
PHASE
SEARCH
HORIZONTAL
VERTICAL
TRIGGER
TRIG D
HELP
ESC
SELECT
RESET
SIMPLE
MODE
POSITION
MATH
MISC
FILE
CLEAR
TRACE
HISTORY
MEASURE
SETUP
PRESET
DISPLAY
CURSOR
CH
1
CH
2
CH
3
CH
4
POWER
COPY IMAGE SAVE
SHIFT
ZOOM
GO/NO-GO
ENHANCED
ACQ
START/STOP
SNAP
SHOT
V DIV
TIME DIV
• To exit the menu during operation, press
ESC
located above the soft keys.
• For a description of the operation using a USB
keyboard or a USB mouse, see section 4.3.
1.
Press
ACQ
. The ACQ menu appears.
2.
Press the
Record Length
soft key.
3.
Turn the
jog shuttle
to select the record length from 1k to 8M.
Explanation
You can set the record length (the number of data points) to be stored to the acquisition
memory.
The selectable maximum record length varies depending on the model.
DL1720E
1 kWord, 10 kWord, 50 kWord, 100 kWord, 250 kWord,
(1 MWord model)
500 kWord (1 MWord)
DL1735E/DL1740E
1 kWord, 10 kWord, 50 kWord, 100 kWord, 250 kWord, 500 kWord,
(2 MWord model)
1 MWord, 2 MWord
DL1740EL
1 kWord, 10 kWord, 50 kWord, 100 kWord, 250 kWord, 500 kWord,
(8 MWord model)
1 MWord, 2 MWord, 4 MWord, (8 MWord)
The value inside the parentheses is selectable only when interleave mode is ON (see the next
section.)
Notes on Setting the Record Length
• The sample rate and display record length vary depending on the T/div setting. For
more details about this relationship, see appendix 1, “Relationship between the Time
Axis, Sample Rate, and Record Length.”
• The maximum record length for box average is half the record length of each model
indicated above.