5-11
IM 701730-01E
Vertical and Horizontal Axes
5
5.7
Using the Preset Function
Procedure
ACTION
DELAY
X - Y
MENU
MENU
PHASE
SEARCH
HORIZONTAL
VERTICAL
TRIGGER
TRIG D
HELP
ESC
SELECT
RESET
SIMPLE
MODE
POSITION
MATH
MISC
FILE
CLEAR
TRACE
HISTORY
MEASURE
SETUP
PRESET
DISPLAY
CURSOR
CH
1
CH
2
CH
3
CH
4
POWER
COPY IMAGE SAVE
SHIFT
ZOOM
GO/NO-GO
ENHANCED
ACQ
START/STOP
SNAP
SHOT
V DIV
TIME DIV
• To exit the menu during operation, press
ESC
located above the soft keys.
• For a description of the operation using a USB
keyboard or a USB mouse, see section 4.3.
Selecting Channels
1.
Press
PRESET
. The PRESET menu appears.
2.
Press the
Select
soft key. The Select menu appears.
3.
Press the
All
or
CH1 to CH4(2)
soft key to select the channel for setting the preset.
Pressing All will select all the channels. The CH3 and CH4 soft keys are not displayed on
the DL1720E.
Selecting the Probe Attenuation/Current-to-Voltage Conversion Ratio
4.
Press the
Probe
soft key. The Probe menu appears.
5.
Press the soft key corresponding to the desired attenuation/current-to-voltage
conversion ratio.