2-46
IM 701210-05E
X-Y Recorder Mode <Section 9.7>
X-Y Waveform Recording
In X-Y waveform recording, the X-Y waveform starts to be displayed when waveform
acquisition is started. If reprint is executed from the Reprint Setup menu after the record
length of waveform (fixed to 1 MW) is acquired or after stopping waveform acquisition,
the X-Y waveform acquired up to the point immediately before the measurement was
stopped is recorded on paper.
The scale of the X-Y waveform recording is 10 mm per division for both X- and Y-axes,
and the waveform recording area is
±
10 divisions (20 divisions). The screen also shows
the X-Y waveform. Recording to paper can also be started automatically when
waveform acquisition is stopped (auto print).
Acquisition Mode during X-Y Recorder Mode
The acquisition mode is automatically set to normal during X-Y record mode. The
acquisition mode cannot be changed with the ACQ key as with the normal operation.
Saving Data in the Internal Memory
In X-Y recorder mode, up to 1 MW of data is automatically saved in the internal memory.
Also, a PDF file of the reprint image can be created.
Functions That Cannot Be Used during X-Y Recorder Mode
When making measurements in X-Y Recorder mode, history memory, GO/NO-GO
determination, dual capture, and waveform computation cannot be used. (The
HISTORY, GO/NO-GO, DUAL CAPTURE, and MATH keys cannot be used.)
In addition, items set with the ACQ key cannot be changed.
Compatibility with the Measured Data
The measured data in X-Y Recorder mode is not compatible with the measured data
when the recorder mode is OFF. The measured/saved data in X-Y Recorder mode
cannot be loaded with the recorder mode turned OFF, and vice versa. The measured
data on the DL750 and the measured data on the DL750P (when the recorder mode is
OFF) are compatible.)
Reprinting on the Built-in Printer (Only during T-Y Recording) <Section 9.8>
After the T-Y recording in Chart Recorder mode is finished, the data stored in the internal
memory can be recalled and reprinted by changing the print conditions.
Creating a PDF File of the Reprint Image (Only during T-Y Recording) <Section 9.9>
After the waveform recording in Chart Recorder mode or X-Y Reorder Mode is finished,
the data stored in the internal memory can be recalled and a PDF file of the reprint
image can be created.
In X-Y Recorder Mode, the most recent 1 MW of data stored in the internal memory are
output to the PDF file. If data exceeding 1 MW is recalled, the waveform displayed on the
screen and the waveform recorded on the printer may differ from the waveform output to
the PDF file.
2.5 Recording in Recorder Mode (Realtime Recording) (DL750P Only)