2-25
IM 701210-05E
Explanation of Functions
2
2.3 Setting the Trigger
• T < T1, T2 < T
A trigger is activated when condition B is met the second time (T), if the time when
condition B is met the second time is outside a specified time range (T1 and T2).
When condition B: CH1=H, Time1=6
µ
s, and Time2=10
µ
s
B is true
CH1
8
µ
s
8
µ
s
B is true B is true
12
µ
s
B is true
Trigger activated here
CH1
B is true
8
µ
s
8
µ
s
B is true B is true
4
µ
s
B is true
rigger activated here
Window Trigger
A certain voltage range (window) is set and a trigger is activated when the trigger source
level enters this voltage range (IN) or exits from this voltage range (OUT)
Width
Trigger is activated
OUT
Trigger is activated
IN
Window
Window
Trigger is activated
Center
Width
Center
IN: Enters the window
OUT: Exits from the window
- : Don’t care
CH1 to CH16, DSP1 to DSP6 (optional)
: Don’t care
: Falling edge
: Rising edge
Edge trigger is used for the logic input.
LogicA/LogicB
Disable: Do not use logic input
Enable: Use logic input