9-19
IM 701210-05E
Recording in Recorder Mode (Realtime Recording) (Only on the DL750P)
9
Selecting the X-Y Waveform to Be Recorded (Displayed)
Select the X-Y waveform to be recorded (displayed) in the range of XY1 to XY4.
Graticule
Type
Select the graticule type from
, OFF, or
.
The graticule setting is linked with the DISPLAY menu > Graticule setting. However, if
OFF is selected, the DISPLAY menu is set to
.
Dark/Light
Set the graticule Dark/Light setting to Light or Dark.
Scale Width of the X- and Y-Axes: Width
Select how to set the scale width of the X- and Y-axes. For details on the format of
the scale lines that are printed, see page 9-10.
1div:
Graticule obtained by dividing the recording area into 10 areas
10mm: mm graticule type
Scale Printing
Select whether to print the scale.
Pointer
Select whether to display the pointer that indicates the most recent recording position on
the screen.
The pointer is only displayed on the screen; it is not printed during X-Y waveform recording.
Font Size: Print Font
Select the size of the printed characters to small or large. The setting applies to all
printed characters.
Comment
A comment string of up to 20 characters can be printed at the bottom right section.
Starting/Stopping X-Y Waveform Recording
When the START/STOP key is pressed, the waveform acquisitions starts, and the
screen shows the X-Y waveform. Pressing the START/STOP key again stops the
waveform acquisition.
If auto print is turned ON, the X-Y waveform recording on the built-in printer starts when
the data acquisition stops. The waveform that is acquired between start and stop is recorded.
The Reprint Setup menu (see section 9.8) can be used to execute X-Y waveform
recording on the built-in printer when auto print is turned OFF.
The most recent 1 MW of data that are acquired during X-Y waveform recording are
saved to the internal memory of the DL750P. The data can be analyzed using
automated measurement of waveform parameters* (see section 11.6), cursor
measurements (excluding the marker cursor. See section 11.5), etc. The data can also
be saved to a storage medium.
*
Area computation on the X-Y waveform is performed only on XY1.
9.7 Recording X-Y Waveforms (X-Y Waveform Recording)