9-7
IM 701210-05E
Recording in Recorder Mode (Realtime Recording) (Only on the DL750P)
9
Explanation
Chart Speed
For T-Y Recording, the chart speed can be selected from below.
10 mm/h, 20 mm/h, 25 mm/h, 50 mm/h, 100 mm/h, 1 mm/min, 2 mm/min, 5 mm/min,
10 mm/min, 20 mm/min, 25 mm/min, 50 mm/min, 100 mm/min, 1 mm/s, 2 mm/s,
5 mm/s, 10 mm/s, and 20 mm/s
You can set the chart speed using the following two methods.
• Menu that appears by selecting RECORDER menu > Chart Speed soft key
• TIME/DIV knob
The top section of the Chart Speed soft key shows the chart speed. The bottom section
shows the horizontal scale (time per division) for T-Y waveform recording.
The chart speed can be changed even while recording is in progress. However, the data
saved in the internal memory is cleared.
Shot Recording
When the specified length of data is recorded, the measurement and recording stop.
The length of the shot recording can be selected from OFF (no shot recording), 20 cm,
50 cm, 1 m, or 2 m.
T-Y Recording to the Built-in Printer: Printer Output
Set whether to carry out T-Y recording to the built-in printer. If set to ON, data is
recorded on the built-in printer and saved to the internal memory. If set to OFF, data is
saved only to the internal memory. The data is not recorded on the built-in printer.
Trigger Mode
In Chart Recorder mode, you can select the trigger mode from Auto, Log, Single, or
Repeat.
Auto
Starts recording to the built-in printer simultaneously with the start of the
measurement. Recording continues until the measurement is stopped.
Log
Starts recording to the built-in printer simultaneously with the start of the
measurement. Measurement and recording stops when the maximum number of
divisions (see page 2-45) of measured values that can be saved is recorded.
Single
Starts recording when the trigger conditions are met after the start of the
measurement. Measurement and recording stops when the maximum number of
divisions (see page 2-45) of measured values that can be saved is recorded.
Repeat
Starts recording when the trigger conditions are met after the start of the
measurement. Recording stops when the shot record length of measured values is
recorded. Then, the DL750P enters the trigger-wait state. The starting/stopping of
the recording is repeated until the measurement is stopped.
For details on the trigger mode in Chart Recorder mode, see page 2-43.
Acquisition Mode
The acquisition mode in Chart Recorder mode is envelope. However, the mode can be
changed to normal mode by using the MISC menu > Others menu.
Envelope and normal modes in Chart Recorder mode are the same as those when
recorder mode is disabled. For details, see page 2-31.
9.3 Setting the Record Conditions of T-Y Waveform Recording