7-21
IM 701210-05E
Acquisition and Display
7
Aborting the Action-on-Stop
15.
Press
START/STOP
. Waveform acquisition is stopped, and action-on-stop is
aborted.
Explanation
Setting the Mode
The specified record length of data is acquired when measurement is started, and the
waveform is displayed. Then, the specified action is executed. In addition, the specified
action is also executed when measurement is stopped. When action-on-stop is enabled,
the trigger mode is set to Log.
Setting the Action
The following five types of actions are available.
Print the Screen Image Data: PRINT
Prints the screen image data on the printer (Printer (built-in printer), USB (USB
printer), Net Print (network printer)) specified by Print to in the PRINT menu.
Save the Screen Image Data: Image
Saves the screen image data to the destination (FD, Zip disk, PC card, internal HD
(optional), SCSI device, or USB storage device) specified in the IMAGE SAVE menu.
Save the Waveform Data: Save to File
Saves the waveform data in binary, ASCII, or floating format to the destination (FD,
Zip disk, PC card, internal HD (optional), SCSI device, or USB storage device)
specified in the FILE menu. The save format is synchronized to data type (see
section 13.7) on the FILE menu.
Buzzer
Sounds a buzzer.
Send Mail
Sends an e-mail message to a specified address (when Ethernet interface option is
installed). Set the number of mail transmissions in the range of 1 to 1000.
For the procedure of setting the address, see section 16.5, “Sending Periodic Mail or
Action Mail.”
Note
If action-on-stop is executed, the data is acquired in Log mode regardless of the trigger mode
setting, and the specified action is carried out. The specified action is also carried out when
the measurement is stopped in the middle of the measurement.
Setting the Number of Actions: Sequence
Sequence
Single:
Performs the action once.
Continue:
Performs the action the number of times specified by ACQ Count.
ACQ Count
2 to 65536:
Repeats the action the specified number of counts.
Infinite:
Repeats the action until waveform acquisition is stopped.
Save to File/Hard Copy/Image Operation
Operates according to the settings in the FILE, PRINT, or IMAGE SAVE menu.
7.8 Setting the Action-on-Stop