7-8
IM 701210-05E
7.4
Acquiring Data Using Box Average
<For a description of this function, refer to page 2-32.>
Procedure
MEASURE CURSOR
RESET
SELECT
CH
ALL CH
MODE
POSITION
SIMPLE/ENHANCED
ACQ
START/STOP
SETUP
DISPLAY
ZOOM
DUAL
CAPTURE
HISTORY
MATH
1
CH
2
CH
3
CH
4
CH
5
CH
6
CH
7
CH
8
CH
9
CH
10
CH
11
CH
12
CH
13
CH
14
CH
15
CH
16
DELAY
MANUAL TRIG
V/DIV
TIME/DIV
TRIGGER
TRIG D
VERTICAL
HORIZONTAL
CAL
X-Y
SEARCH
7
DSP 1
8
9
DSP 2
6
5
4
1
0
2
3
ENTER
m
DSP 6
DSP 5
LOGIC A
LOGIC B
EVENT
DSP 3
DSP 4
EXP
GO/NO-GO
ACTION
FILE
SHIFT
1.
Press
ACQ
.
2.
Press the
Mode
soft key. The mode selection menu appears.
3.
Press the
Box Average
soft key.
Explanation
This function is valid only on the 701250 (HS10M12) and the 701255 (NONISO_10M12).
At sample rates that allow box averaging, data is normally sampled at 10 MS/s and the
sampled data is decimated according to the T/div setting and stored to the acquisition
memory. The internal sample rate is the same as the envelope mode.
For details, see appendix 1, “Relationship between the Time Axis Setting, Sample Rate
and Record Length.”
When you use the box averaging function, the operation is different. Moving average is
computed on a given number of data points and the result is sampled and stored to the
acquisition memory.
Time
Time
1
2
20
a
21
Input signal
When set to 50 kS/s
Box averaged data
b
. . . . . . .
36
. . . . . . .
. . .
Voltage
Voltage