6-5
IM 701210-05E
Triggering
6
6.3
Setting the Trigger Delay
<For a description of this function, refer to page 2-28.>
Procedure
MEASURE CURSOR
RESET
SELECT
CH
ALL CH
MODE
POSITION
SIMPLE/ENHANCED
ACQ
START/STOP
SETUP
DISPLAY
ZOOM
DUAL
CAPTURE
HISTORY
MATH
1
CH
2
CH
3
CH
4
CH
5
CH
6
CH
7
CH
8
CH
9
CH
10
CH
11
CH
12
CH
13
CH
14
CH
15
CH
16
DELAY
MANUAL TRIG
V/DIV
TIME/DIV
TRIGGER
TRIG D
VERTICAL
HORIZONTAL
CAL
X-Y
SEARCH
7
DSP 1
8
9
DSP 2
6
5
4
1
0
2
3
ENTER
m
DSP 6
DSP 5
LOGIC A
LOGIC B
EVENT
DSP 3
DSP 4
EXP
GO/NO-GO
ACTION
FILE
SHIFT
1.
Press
SHIFT+POSITION
.
2.
Turn the
jog shuttle
to set the trigger delay.
Press
RESET
to reset the value corresponding to the selected unit to 0.0
µ
s.
Explanation
Although the display usually shows the waveform before and after the trigger point, it is
possible to display the acquired waveform after a fixed time period elapses using the
delay function.
Selectable Range of Trigger Delay
0 to 10 s
(Minimum resolution is 100 ns.)
Delay
Trigger point
T (Trigger position)
Notes for Setting the Trigger Delay
• When T/div is changed, the trigger delay remains unchanged.