5.18 Setting the Acceleration Measurement
<For a description of this function, refer to page 2-15.>
CAUTION
Before connecting the acceleration sensor, check that the bias current to the
sensor is OFF (see step 8 below). Otherwise, damage to the internal circuitry of
the acceleration sensors may result.
Procedure
MEASURE CURSOR
RESET
SELECT
CH
ALL CH
MODE
POSITION
SIMPLE/ENHANCED
ACQ
START/STOP
SETUP
DISPLAY
ZOOM
DUAL
CAPTURE
HISTORY
MATH
1
CH
2
CH
3
CH
4
CH
5
CH
6
CH
7
CH
8
CH
9
CH
10
CH
11
CH
12
CH
13
CH
14
CH
15
CH
16
DELAY
MANUAL TRIG
V/DIV
TIME/DIV
TRIGGER
TRIG D
VERTICAL
HORIZONTAL
CAL
X-Y
SEARCH
7
DSP 1
8
9
DSP 2
6
5
4
1
0
2
3
ENTER
m
DSP 6
DSP 5
LOGIC A
LOGIC B
EVENT
DSP 3
DSP 4
EXP
GO/NO-GO
ACTION
FILE
SHIFT
1.
Press one of the keys from
CH1
to
CH16
to select the desired channel (select a
channel with the 701275(ACCL/VOLT) installed).
2.
Press the
Coupling
soft key followed by the
ACCL
soft key.
Setting the Gain
3.
Press the
Gain
soft key and select a value in the range of
×
0.1 to
×
100.
Setting the Sensitivity
4.
Press the
Next 1/2
soft key.
5.
Press the
Sensitivity
soft key.
6.
Turn the
jog shuttle
and set the sensitivity of the acceleration sensor in the
range of 0.10 mV/Unit to 2000.00 mV/Unit.
Setting the Unit
The default value is m/s
2
. Change the unit as necessary.
7.
Press the
Unit
soft key to display a software keyboard, and enter the unit
according to the procedure given in section 4.2.
Connecting the Acceleration Sensor
8. Press the
Bias
soft key to select OFF.
9. Connect the acceleration sensor according to the procedures given in section 3.12.
Setting the Bias Current Supply
10.
Press the
Bias
soft key to select ON.