Recorder Output (Analog Output)
Recorder output is connected to observation equipment (recorder, oscillo-
scope, etc.), and records of illuminance variation or a light source waveform
can be observed.
Specifications: 999mV
±
5% (range fixed for full scale of each range) Load
resistance of 100 k
Ω
or more
1. Prepare a coated single-core shielded wire and solder it to the attached
recorder output plug as shown in Figure 10. Connect another shielded
wire to the input connector of the observation equipment.
Shielded wire
Soldered on + side
Soldered on - side
Plug cover
Figure 10. Shielded Wire Connection Diagram
2. Turn on the power to the illuminance meter and insert the plug into the
recorder output connector of the light detector. The display [
REC
, R-
H] appears, and the range is fixed (manual range).
3. Set the response selector switch to FAST.
4. The output voltage is 1 mV for one numeral of the least significant digit
of the effective display (3 digits). Confirm the range according to the
illuminance-measuring range. Use the
RANGE
key for range setting.
Note: Refer to “Range Hold” for
RANGE
key operation.
5. Cover the light detector with the cap to adjust the zero level of the
observation equipment.
6. Remove the cap from the light detector to start observation.
· The application of this accessory could affect the immunity performance of
the product and such items should be used with this consideration.
· Interconnection with other apparatus is not recommended.
Note