background image

User's

Manual

IM3E-2006.2

Yokogawa Meters & Instruments Corporation
International Sales Dept.

Tachihi Bld. No.2, 6-1-3, Sakaecho, Tachikawa-shi,Tokyo 190-8586 Japan

Phone: 81-42-534-1413, Facsimile: 81-42-534-1426

YOKOGAWA CORPORATION OF AMERICA (U.S.A.)

Phone: 1-770-253-7000  Facsimile: 1-770-251-2088

YOKOGAWA EUROPE B. V. (THE NETHERLANDS)

Phone: 31-334-64-1611  Facsimile: 31-334-64-1610

YOKOGAWA ENGINEERING ASIA PTE. LTD. (SINGAPORE)

Phone: 65-6241-9933  Facsimile: 65-6241-2606

YOKOGAWA AMERICA DO SUL S. A. (BRAZIL)

Phone: 55-11-5681-2400  Facsimile: 55-11-5681-1274

YOKOGAWA MEASURING INSTRUMENTS KOREA CORPORATION (KOREA)

Phone: 82-2-551-0660 to -0664  Facsimile: 82-2-551-0665

YOKOGAWA AUSTRALIA PTY. LTD. (AUSTRALIA)

Phone: 61-2-9805-0699  Facsimile: 61-2-9888-1844

YOKOGAWA INDIA LTD. (INDIA)

Phone: 91-80-4158-6000  Facsimile: 91-80-2852-1441

YOKOGAWA SHANGHAI TRADING CO., LTD. (CHINA)

Phone: 86-21-6880-8107  Facsimile: 86-21-6880-4987

YOKOGAWA MIDDLE EAST E. C. (BAHRAIN)

Phone: 973-358100  Facsimile: 973-336100

LTD. YOKOGAWA ELECTRIC (RUSSIAN FEDERATION)

Phone: 7-095-737-7868  Facsimile: 7-095-737-7869

Printed in Japan

Model 3226

Universal Leakage Current Tester

IM 3226-E

Nov. 2007 5th (KP)

Read this user’s manual before using the instrument in order to fully and correctly utilize 

all its functions.

   Model: 322610 Universal Leakage Current Tester

 

322710 Test Box

GENERAL

Model 3226

 is a special milliammeter for measuring leakage current of electrical appli-

ances. The input resistance can be selected from among 1, 1.5 and 2 KΩ for easy measure

-

ment.

Model 3226

 is usable for product inspection on production lines and for experimental 

measurement in research and design activities. 

Model 3227

 Test Box is available for the 

convenience of measurement.

Model 3226

 has the following features:

(1)  4-way function: AC current, DC current, DC + AC current and AC voltage are mea-

surable.

(2)  High accuracy: ±2.5% of full scale.

(3)  High sensitivity: 0.1 mA for full scale.

(4)  TAUT BAND system meter free of friction and strong against vibration and shock.

(5)  Overload protective circuit to prevent the meter coil from burning and the meter 

pointer from bending

(6)  Shielded case to eliminate effect of external high-frequency electric field.

(7)  Compact and lightweight for easy portable use.

SAFETY PRECAUTIONS

When operating the instrument, be sure to observe the cautionary notes given below to 

ensure correct and safe use of the instrument. Yokogawa is by no means liable for any 

damage resulting from use of the instrument in contradiction to these cautionary notes.

The following safety symbols are used on the instrument and in this manual.

Warning

  Indicates a hazard that may result in the loss of life or serious 

injury of the user unless the described instruction is abided by.

Caution

  Indicates a hazard that may result in an injury to the user and/or 

physical damage to the product or other equipment unless the 

described instruction is abided by.

Danger!

 Handle with Care. 

This symbol indicates that the operator must refer to an explana-

tion in the User’s Manual in order to avoid risk of injury or death of 

personnel or damage to the instrument.

    This symbol indicates AC voltage/current.

    This symbol indicates a Fuse.

    This symbol indicates ground(earth).

Warning

To avoid injury, death of personnel, carefully observe and follow the warnings listed below:

 Measurement

- Do not operate the instrument over 250 V of strong electric circuit.

- Do not touch the Input (Output) terminals, when measuring voltage.

 Measuring lead

- Use the lead supplied by Yokogawa for the instrument concerned.

- Do not use a deteriorated or damaged lead.

- Do not attach/detach the lead to /from the instrument prior to releasing it from the 

measured object.

 Protection

- Be sure to use the designated fuse (Rating: voltage, current and type) to prevent fire. 

- If there are any cracks or other damage in the case because of being dropped or 

struck, the instrument may not be safety insulated. Do not use the instrument before 

any remedial measures are taken.

 Replacement of batteries or fuse

- Prior to detaching the cover for replacing the batteries or fuse, release the lead from 

the measured object and turn off the switch.

 Operating Environment

- Do not operate the instrument in a flammable or explosive gas atmosphere.

- Do not operate the instrument if there is condensation on it.

 Disassembly

- Do not attempt to disassemble the instrument.

Caution

To avoid injury of personnel or damage to the instrument, carefully observe and follow 

the cautions listed below:

 Measurement

  Do not apply a voltage and current over the allowable limits between the terminals.

 Selector switch

  Do not switch the Measuring range selector switch during measurements.

 Batteries

- Do not use different types of batteries together or new and old batteries together.

- If the instrument is not used for a long period, store the instrument with the batteries 

removed.

  Otherwise, any leakage from the batteries may damage the instrument.

Note

Model 3226

 has AC + DCmA range. This is used mainly for the instrument which is 

operated by DC after rectifing, AC to DC. Therefore, measured value on this range does 

not indicate RMS value, but the sum of DC and AC components directly.

BLOCK DIAGRAM

Overload

protective

circuit

Battery

check

circuit

Range

selector

circuit

Input

resistance

selector

circuit

Operating

amplifier

Moving-coil

type DC

milliammeter

200 µA F.S.

Internal

resistance: 121 Ω 

 PARTS IDENTIFICATION AND FUNCTION

B

ACV

B

0

2

4

6

90

180

120

240

150

300

8

10

AC+DCmA

ACmA

DCmA

ACV

10

1kΩ

1.5kΩ

2kΩ

INPUT

RESISTANCE

10

10

1

1

1

0.1

0.1

INPUT

0.1

150

300

BATT

ON

OFF

6. Scale plate

7. Carrying

handle

8. Meter cover

fixing screw

9. Meter cover

10. Measuring

range selector

switch

11. Input

terminals

12. Case

1. B mark

2. Spear 

pointer

3. Zero

adjuster

screw

4. Input

resistance

switch

5. Power

on-off/battery 

check switch

Fig. 1 Parts Identification

1.  B mark:

 

The battery is usable if the meter pointer is within this range when the BATT/ON/OFF 

switch is set to the BATT position.

2.  Spear pointer

3.  Zero adjuster screw:

 To be turned with a screwdriver for readjusting zero indication of the pointer.

4.  lnput resistance switch:

 Selects one of the three input resistances according to the applicable standard.

5.  Power on-off/battery check switch

6.  Scale plate:

 

The scale is double graduated in mA and AC V.

7.  Carrying handle

8.  Meter cover fixing screw

9.  Meter cover

10. Measuring range selector switch:

 To be set to an appropriate position according to the measuring item and 

range.

11. lnput terminals:

 

Accessory H-lead is to be connected to the “INPUT” terminal, and L-lead to the “ ” 

(ground) terminal.

12. Case

A3