IM 2768-01E
― 9 ―
5. PRECAUTIONS
(1) A measurement with high accuracy must be in a room where the temperature is stable to
18 to 25°C and where humidity is low.
After completion of the measurement, be sure to set the switches of
POWER SUPPLY, GA and BA to OFF.
(2) In measurement, be sure to depress GA switch after BA switch has been depressed.
If this order is reversed, the galvanometer pointer will misleadingly deflect for
the measuring dial operation inductance of the resistor under test
when BA switch is depressed.
(3) The resistance of the lead wires which connect the resistance under test (Rx) to
the Rx terminals, is included in the measured value.
Therefore, in measuring the resistance lower than 100 Ω,
the influence of the lead wire resistance must be eliminated by measuring
the resistor which is correctly calibrated, or by measuring the resistance of the lead wires.