11-4
IM 2553A-01EN
11-4
IM 2553A-01EN
:OUTPut[:STATe]
Function
Sets or queries the output’s on/off state.
Syntax
:OUTPut[:STATe] <Boolean>
:OUTPut?
ON|1: Enabled
OFF|0: Disabled
Example
:OUTPUT ON
:OUTPUT? -> :OUTPUT 1
Description If the source function is RJTemp and you set
the output’s on/off state, a “221: Setting conflict”
error will occur. Set the source function with the
:SOURce:FUNCtion command.
:OUTPut:POLarity
Function
Sets or queries the polarity.
Syntax
:OUTPut:POLarity {NORMal|INVerted}
:OUTPut:POLarity?
NORMal: Positive polarity
INVerted: Negative polarity
Example
:OUTPUT:POLARITY NORMAL
:OUTPUT:POLARITY?
-> :OUTPUT:POLARITY NORMAL
Description • If the source function is RJTemp and you set
the polarity, a “221: Setting conflict” error will
occur.
NORMal is returned as a response
to a query. Set the source function with the
:SOURce:FUNCtion command.
• If the output setting goes outside the selectable
range of the source function when you change
the polarity, the output will turn off. In this
condition, the output cannot be turned on.
11.3 OUTPut Group
The commands in this group deal with the output on/off setting and polarity setting. The commands correspond to the
front panel OUTPUT and POLARITY switches.