13-4
IM 2553A-01EN
13.3 Error Code Descriptions and Corrective
Actions
This section explains error codes that may appear on the front panel.
Error Codes That Appear at Power-on or during the Self-test
If any of the following error codes appears, servicing is required.
Code
Description
E.911 to E.914
An internal circuit error was detected.
Warning Codes
Code Description
Corrective Action
Refer To
E.032 Excessive input was detected during resistance
sourcing.
Check the connection conditions or load.
3.5, 3.6
E.033 An oscillation error was detected in the voltage
or current amplifier.
Check the connection conditions or load.
3.5, 3.6
E.034 An overvoltage load was detected.
Check the connection conditions or load.
3.5, 3.6
E.035 An overcurrent load was detected.
Check the connection conditions or load.
3.5, 3.6
E.040 An error was detected in the user definition file. Check the syntax.
Appendix 4