![YEO-KAL 801C Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/yeo-kal/801c/801c_user-manual_3283168013.webp)
13
Upload Data from S618 to PC
Clear Store Memory (1)
From the
main menu
select
logger settings
press the
ENTER
and then selecting
Clear
801B (or 620)
Memory.
Press
ENT
and the operator will proceed and clear memory.
Calibration Constants to PC (2)
The calibration constants can be uploaded to a PC for quality assurance reports. The information
shows
Date, Time
and
calibration coefficients
.
Store Data out (3)
The reader unit has two separate memories: the store memory and the logger memory. The logger
memory can hold up to 16384 readings and the store memory can hold up to 4096 readings. Once
stored, the data can then be down loaded into a computer via Bluetooth or data cable.
The store memory is used to hold spot measurements. To store a reading while in the field simply
press the
STORE
key. The data will be down loaded in the order in which it is stored so keep a
record of the site at which each measurement was made so that the data can be easily analysed.
Logged Data Out (4)
Data stored in the YK801B can be uploaded to a computer by Bluetooth or dada cable and then
stored, graphed or printed out using most popular applications. When the data is uploaded, it
includes a header, as shown in figure 8 below, indicating whether the data is from the logger or the
store memory.
Figure 8 Data header as:
Total samples while Logging with Unit 618 51
ID, Time , Temp (C), D.O. (%sat),Latitude, Longitude
ID: 668, Start Time: 18:38:25 18/07/21
1 18/07/21 18:38:25 15.34 100.6 000 00.0000,000 00.0000
2 18/07/21 19:08:25 15.21 100.7
3 18/07/21 19:38:25 15.08 100.7
4 18/07/21 20:08:25 14.95 100.8
5 18/07/21 20:38:25 14.83 100.8