TUNICS T100S-HP User Manual
37
6. Calibrating the Product
6.1
Performing Power Calibration
Subject
The power calibration procedure calibrates the laser output power. The procedure
requires the use of an optical power meter.
Procedure
1.
Connect the laser output of the product to a power meter, (see section
External Optical Devices to the product, p. 22
).
2.
Enable the laser output (see section
Enabling/Disabling the Laser Output, p. 26
)
.
3.
Press the
2nd
+
P
function keys to activate the power calibration procedure.
The tunable laser moves to a short wavelength within the tuning range and displays
the power level corresponding to the current calibration data.
4.
Measure the power level at the first wavelength on the power meter.
5.
Enter the reading at the prompt
P =
.
The tunable laser moves to a long wavelength within the tuning range and displays
the power level corresponding to the current calibration data.
6.
Measure the power level at the second wavelength.
7.
Enter the reading at the prompt
P =
.
The power calibration is now complete and the instructions reference the new power
calibration data.
This procedure corrects the factory-set calibration data.
•
Do not perform this procedure unless you actually intend to correct this
data.
•
Do not perform these procedures unless you have calibrated sources.
CAUTION
Содержание TUNICS T100S-HP
Страница 2: ......
Страница 6: ...About This Manual 6 TUNICS T100S HP User Manual...
Страница 24: ...Installing and Connecting the Product 24 TUNICS T100S HP User Manual...
Страница 36: ...Changing the Optical Output Power and Diode Current 36 TUNICS T100S HP User Manual...
Страница 50: ...Using the Auxiliary Inputs and Outputs 50 TUNICS T100S HP User Manual...
Страница 54: ...Verifying Performances 54 TUNICS T100S HP User Manual...
Страница 60: ...Certification and Compliance 60 TUNICS T100S HP User Manual...
Страница 66: ...Index 66 TUNICS T100S HP User Manual...