background image

Changing the Optical Output Power and Diode Current

34

TUNICS T100S-HP User Manual

The active mode is 

constant-power

: the laser-diode current is controlled to ensure 

an optical power output equal to the 

P

 parameter.

By pressing the 

P

 function key, the system switches automatically to the constant-

power mode.
In this mode, the required power cannot be obtained, even with the maximum 

allowable current. In this case:

The current is limited to its maximum value.

The optical power is lower than the user-set power value.

The 

Lim

 message is displayed to the right of the current and power values.

To activate the constant-current mode, press the APC function key to turn 

off

 the APC 

key LED:
The active mode is 

constant-current

: the diode current level is determined by the 

I

 

parameter.
By pressing the 

I

 function key, the system switches automatically to the constant-

current mode.

5.3

Disabling/Enabling Active Cavity Control

Subject

The TUNICS T100S-HP is equipped with on-board active cavity control.
When you turn on the TUNICS T100S-HP after the initialization routine is complete, the 

system switches automatically to APC (constant power) mode with active cavity control 

(the APC LED is on). 

Active Cavity Control Requirements

Active cavity control is enabled if the following requirements are both met:

APC (constant power) mode (APC LED on)

Power > 0.8 mW

System Status

System status is summarized in the following table:

Table 11: Active Cavity Control - System Status

Procedure

To disable active cavity control, press the

2nd

+

APC

key combination.

The 

APC

 LED blinks. 

If you press the 

APC

 key and the APC LED goes off, this indicates that the system has 

switched in constant current mode. When entering constant current mode, the active 

cavity control becomes disabled, and the system restores the operating parameters 

that were in effect at the time the TUNICS T100S-HP last left the constant current 

mode.

To enable active cavity control, press the 

2nd

+

APC

 command keys again.

If the APC LED starts blinking while the TUNICS T100S-HP is in APC mode with active 

cavity control on, this may indicate that the laser power output (P) is set to a value 

lower than the 0.8 mW threshold.

APC LED

Constant Power Mode

Active Cavity Control

On

Yes

Yes

Blinking

Yes

No

Off

No

No

Содержание TUNICS T100S-HP

Страница 1: ...3 0 1 5 5 4 5 4 6 3 5 www yenista com TUNICS T100S HP High Power Tunable Laser User Manual TUNICS T100SHP_UM_6 04v1 9...

Страница 2: ......

Страница 3: ...hat instrument Yenista does not warrant that the operation of the instrument software or firmware will be uninterrupted or error free The foregoing warranty shall not apply to defects resulting from i...

Страница 4: ...ng 1 Room 1016 Shanghai 200082 CHINA Phone 86 21 5160 7615 Sales Sales Americas Email sales am yenista com Phone 1 805 367 4075 Sales China Email sales china yenista com Phone 86 21 5160 7615 Sales Eu...

Страница 5: ...IEEE 488 1 interfaces used to operate the TUNICS T100S HP in the remote modes Date 26 April 2016 Manual Reference TUNICS T100SHP_UM_6 04v1 9 Typographical Conventions Symbols bold Identifies interfac...

Страница 6: ...About This Manual 6 TUNICS T100S HP User Manual...

Страница 7: ...the Laser 25 3 2 Enabling Disabling the Laser Output 26 3 3 Shutting Down the Product 26 4 Setting Up the Emission Wavelength or Frequency 27 4 1 Changing the Emission Wavelength or Frequency 27 4 2 S...

Страница 8: ...dulation 48 8 2 Output Signals 49 8 2 1 Available TTL Output Signals 49 8 2 2 Monitoring Outputs 49 8 2 3 Activating Deactivating the Output as a Trigger 49 9 Verifying Performances 51 9 1 Preparing t...

Страница 9: ...ical Safety Label This label indicates an injury hazard It appears on a location that requires special instructions for proper use you must consult the documentation before using the element of the pr...

Страница 10: ...oes not apply more than 265 Volts RMS between the supply conductors or between either of the supply conductors and the ground To avoid the possibility of injury make sure the socket outlet in which th...

Страница 11: ...g with remarkable wavelength accuracy Internal wavelength referencing The TUNICS T100S HP features its own internal referencing system to ensure highly accurate and repeatable measurements Active wave...

Страница 12: ...escription Essential Models Extended Range Models O CL O ES SCL CLU Wavelength range nm 1260 1360 1500 1630 1240 1380 1350 1510 1440 1640 1500 1680 Output power Over full wavelength range 10 dBm 8 dBm...

Страница 13: ...at constant temperature 4 Measured at 100 MHz 5 O and CL at 10 dBm Others at 8 dBm 40 pm all at 0 dBm 6 Validated at 0 and 10 dBm for essential and 0 and 8 dBm for extended range models 7 Over 100 wav...

Страница 14: ...y 80 non condensing Altitude maximum operating 2000 meters Essential Models Extended Range models O CL O ES SCL CLU Laser Diode Current Level I See the Acceptance Test Report delivered with the produc...

Страница 15: ...s 1 3 1 Description of the Front Panel Figure 1 Front Panel DATA Area Description The DATA area displays the three main operating parameters Emission wavelength or optical frequency f Output power P L...

Страница 16: ...and Constant Current Modes p 33 STEP Changes the wavelength increment STEP value see section Setting a Predefined Increment Decrement Step p 28 It is also possible to increase or decrease the wavelen...

Страница 17: ...icrometer screw see section Turning On Off Backlash Suppression of the Micrometer Screw p 32 2 To display the software release Remote To change the GPIB address and to activate deactivate the RS 232 r...

Страница 18: ...ng I BNC connector for current monitoring TTL Outputs Mod Sync BNC connector for modulation sync TTL output Mod Out BNC connector for modulation output TTL output For a detailed description on particu...

Страница 19: ...ty of laser products see section Certification and Compliance p 59 This means that it is safe under all conditions of normal use except when viewed with optical instruments such as microscopes and tel...

Страница 20: ...ction Certification and Compliance p 59 EFUP Label RoHS China The product has an environment friendly use period of 40 years See section Certification and Compliance p 59 Warranty Seal The product cov...

Страница 21: ...the rear panel see Figure 2 p 17 make sure the power switch is set to 0 7 On the front panel in the OUTPUT area make sure the key lock switch is in STBY position before applying power 8 Place the TUNI...

Страница 22: ...ty requirements The TUNICS T100S HP has a chassis connected to ground via the power supply cable A protective ground connection by way of the grounding conductor in the power cord is essential for saf...

Страница 23: ...ts the following pin assignment Pins 1 2 Shorted Pins 6 7 Connect to external switch Figure 4 Remote Interlock SUBD 9 Connector Opening the interlock switch disables the laser output To turn on the la...

Страница 24: ...Installing and Connecting the Product 24 TUNICS T100S HP User Manual...

Страница 25: ...when turning on the system 4 To skip the self calibration step skip Referencing while the message Initializing is displayed press the 0 key Once the initialization routine is complete the laser output...

Страница 26: ...ars on the third display 2 On the front panel in the OUTPUT area turn the key lock switch counterclockwise to the STBY position The message Parking displays and the optical head is automatically moved...

Страница 27: ...the wanted value is reached turn right to increase the value or left to decrease the value To modify the basic increment decrement applied by the rotary knob press the and arrow keys to set the scrol...

Страница 28: ...age Lim appears 4 In the SETUP area press the ENTER key 5 Increase or decrease the wavelength by pressing the or arrow keys in the SETUP area 4 3 Using the Fine Scanning Mode Fine Tuning Subject The F...

Страница 29: ...at you set During each scan interval the TUNICS T100S HP laser broadcast at a set wavelength for the duration specified by the Pause Timer parameter that you set Before Starting Before activating the...

Страница 30: ...u to enter the scan STEP while the current setting is displayed 5 Enter the scan STEP and press Enter To know the valid range for scan STEP values see section Valid Range for Parameter Settings p 14 T...

Страница 31: ...displayed 4 Enter 0 as scan STEP parameter and press Enter The DATA area display prompts you to enter the Pause Timer The PAUSE TIMER is the dwell time at the upper wavelength of the sweep range befo...

Страница 32: ...m to match the nearest operational value i e 67 nm s 3 Press Enter to confirm the change 4 6 Turning On Off Backlash Suppression of the Micrometer Screw Subject The TUNICS T100S HP firmware systematic...

Страница 33: ...required output power may not be reached even at the maximum allowable current If this occurs the TUNICS T100S HP automatically limits the current to its maximum value and the message Lim appears at t...

Страница 34: ...100S HP after the initialization routine is complete the system switches automatically to APC constant power mode with active cavity control the APC LED is on Active Cavity Control Requirements Active...

Страница 35: ...nd the current parameter value is displayed When the sign flashes you can modify the value 2 Do one of the following Using the numeric keypad enter the wanted value If you make a mistake while enterin...

Страница 36: ...Changing the Optical Output Power and Diode Current 36 TUNICS T100S HP User Manual...

Страница 37: ...length within the tuning range and displays the power level corresponding to the current calibration data 4 Measure the power level at the first wavelength on the power meter 5 Enter the reading at th...

Страница 38: ...to 0 mA and 0 mW respectively Therefore as long as the system is in remote interlock mode it cannot perform any self calibration or internal referencing Procedure 1 Press the 2nd I function keys to s...

Страница 39: ...iguration requirements Serial port command set based on the SCPI Standard Commands for Programming Instruments syntax Serial interface configuration RS 232 ON Baud rate 9600 Handshaking None Depending...

Страница 40: ...re Figure 6 Instrument Setup for Wavelength Monitoring 2 Connect a communication link between the TUNICS T100S HP tunable laser source and the wavemeter via the RS 232 C interface make sure the RS 232...

Страница 41: ...s implemented The wavemeter actual reading loops back to the tunable laser An iteration process self aligns the tunable laser Figure 7 Wavelength Monitoring Scheme Depending on the performance of the...

Страница 42: ...at the front panel indicator until the output wavelength has reached the target value and has stabilized within wavemeter accuracy To compensate for possible disturbance such as accidental vibrations...

Страница 43: ...signals in the APC constant power mode from 30 kHz to 8 MHz for sinusoidal signals from 30 kHz to 1 MHz for TTL signals Modulation of the diode current is obtained by applying a signal anywhere betwe...

Страница 44: ...HP R is the 50 resistor of the LF modulation circuit With a maximum allowed voltage input of 5 volts the maximum input current you can add to the TUNICS T100S HP laser diode is 100 mA Resulting LF Mo...

Страница 45: ...y of the modulation signal is well above the intrinsic frequency 1 kHz of the APC loop the average power is monitored and remains constant Modulation is possible in the APC mode at frequencies above 3...

Страница 46: ...he relationship between the voltage at the HF Mod input and the current added to the diode is given by IMPORTANT Before implementing external modulation via HF Mod input connector make sure the system...

Страница 47: ...compute the total current using the following formula Where Idiode is the total current through the diode Ibias is the bias current Vin is the voltage applied to the HF Mod input Rinternal is the 50 T...

Страница 48: ...e TUNICS T100S HP has a wavelength fine tuning capability that can be used to adjust the wavelength very precisely or to serve the optical frequency This function is available via the FSC connector on...

Страница 49: ...2 3 Activating Deactivating the Output as a Trigger Subject You can switch the output to a trigger TTL that is high when the TUNICS T100S HP is scanning only in positive direction ignoring return and...

Страница 50: ...Using the Auxiliary Inputs and Outputs 50 TUNICS T100S HP User Manual...

Страница 51: ...ibed in section Initializing the Laser p 25 3 Connect the laser output as described in section Connecting External Optical Devices to the product p 22 4 Enable the laser output as described in section...

Страница 52: ...y that the measured wavelength is no greater than 0 04 nm away from each wavelength setting 9 3 Verifying Wavelength Range and Output Power Subject The following procedure verifies that the laser outp...

Страница 53: ...any residual emissions is lower than 45 dB down from the peak level 9 5 Verifying Wavelength and Power Calibration Subject During the course of the performance verification it is possible that the set...

Страница 54: ...Verifying Performances 54 TUNICS T100S HP User Manual...

Страница 55: ...the connectors to remove any loose particles or moisture 3 Moisten a clean optical swab with isopropyl alcohol and lightly swab the surfaces of the connectors To avoid personal injury do not operate t...

Страница 56: ...se Before Starting Unplug the power cord Make sure you have the appropriate power fuse see section Technical Specifications p 12 Procedure 1 Insert a small flat blade screwdriver into the notch just i...

Страница 57: ...EEPROM error Turn Power off Electrical problem Turn off the system and contact a service engineer immediately Referencing error Press Enter The system s internal wavelength referencing operation has f...

Страница 58: ...rror and Warning Messages DEFAULT Laser Head Overheating The temperature of the room is too high the laser is disabled Switch off the product and make sure the temperature room is lower than 30 C befo...

Страница 59: ...ctive The product is marked with the WEEE symbol for recycling This means that you must not discard the product with domestic household waste At the end of the product s useful life please contact you...

Страница 60: ...Certification and Compliance 60 TUNICS T100S HP User Manual...

Страница 61: ...ions 14 Table 5 Valid Range for Parameter Settings 14 Table 6 Available Accessories and Replaceable Parts 14 Table 7 DATA Area Function Keys 15 Table 8 MODE Area Function Keys 17 Table 9 SETUP Area Ke...

Страница 62: ...ote Interlock SUBD 9 Connector 23 Figure 5 Wavelength scanning process 29 Figure 6 Instrument Setup for Wavelength Monitoring 40 Figure 7 Wavelength Monitoring Scheme 41 Figure 8 Low frequency modulat...

Страница 63: ...ive 59 Cleaning 55 Coherence Control Enable Disable 32 Input 48 Compliance 59 Connections to instruments 22 Constant current mode 33 Constant power mode 33 Current change 35 Current Saturation 47 D DA...

Страница 64: ...nm GHz key 15 Numeric keypad 17 O Operating mode Constant current mode 16 Constant power mode 16 Manual mode 17 Optical connectors 55 OUTPUT area 17 Output power Change 33 Verification 52 Outputs 49...

Страница 65: ...fications 12 Test preparation 51 Trigger 49 TTL outputs 18 49 U Units 15 V Valid range 14 Verification 51 W Warning messages 57 Wavelength Change 27 Modulation 48 Monitoring 39 Range verification 52 S...

Страница 66: ...Index 66 TUNICS T100S HP User Manual...

Отзывы: