Installing and Connecting the Product
TUNICS T100S-HP User Manual
23
2.4
Installing the Remote Interlock
Subject
The TUNICS T100S-HP is equipped with a Remote Interlock connector at the rear-panel.
You can connect an external remote interlock switch to the Remote Interlock SUBD-9
connector located on the rear panel.
Remote interlock sets the system current and power to 0 mA and 0 mW, respectively.
Before Starting
Procedure
•
To connect the remote interlock switch, use a connector that meets the following pin
assignment:
•
Pins 1-2: Shorted
•
Pins 6-7: Connect to external switch
Figure 4: Remote Interlock SUBD-9 Connector
•
Opening the interlock switch disables the laser output.
•
To turn on the laser output, close the interlock switch and press the
Enable
key.
When the system resumes from remote interlock mode into active state, we suggest
that you run a new system auto-calibration procedure (see section
Internal Wavelength Referencing, p. 38
) since the present system calibration data is
still based on the previous system operating conditions.
IMPORTANT
Make sure that the external switch is electrically isolated from other
circuits, including earth ground.
External switch
Содержание TUNICS T100S-HP
Страница 2: ......
Страница 6: ...About This Manual 6 TUNICS T100S HP User Manual...
Страница 24: ...Installing and Connecting the Product 24 TUNICS T100S HP User Manual...
Страница 36: ...Changing the Optical Output Power and Diode Current 36 TUNICS T100S HP User Manual...
Страница 50: ...Using the Auxiliary Inputs and Outputs 50 TUNICS T100S HP User Manual...
Страница 54: ...Verifying Performances 54 TUNICS T100S HP User Manual...
Страница 60: ...Certification and Compliance 60 TUNICS T100S HP User Manual...
Страница 66: ...Index 66 TUNICS T100S HP User Manual...