52
OK
13.1.2
Menu
→
Diagnostic
→
Gas probe
Tightness test of the probe.
15/06/22 10:00
GAS PROBE
Close the flue
gas probe
Push OK to
continue
15/06/22 10:00
GAS PROBE
Close the flue
gas probe
Push OK to
continue
15/06/22 10:00
GAS PROBE
Calibration
15/06/22 10:00
GAS PROBE
Probe Test
15/06/22 10:00
GAS PROBE
Calibration
Probe Test
Result: Tight
→
→
KEY
FUNCTION
Returns to the previous screen.
Starts the test to check the tightness of the gas sampling probe.
OK
ESC
Connect the flue gas sampling probe and filter unit
assembly to the instrument;
Fully insert the black rubber cap (68624) on the gas probe
tip, as shown in the following picture:
Provided black rubber cap
Results:
Tightness
:
The system is OK
Leak
:
Make sure that the probe is connected to the input P
-
or P+, check the seals of the pneumatic
connections and/or the seal of the condensation trap and check that the test cap is correctly inserted
on the tip of the probe.
WARNING: a damaged probe tip may impair the test.
Error
:
It is not possible to perform the test because the sensor is not calibrated.