11
2. PRE-OPERATION PROCEDURES
Installation
7.
Install the ion exchange resin cartridge.
●
Turn power off and close water tap before installing or replacing the ion exchange resin
cartridge.
●
Install procedure:
(1) Remove rubber cap from ion
exchange resin cartridge.
(2) Install ion exchange resin cartridge
securely
into
the
resin
drum
installation plate, inside the WE200
unit. (See Figure 1.)
(3) Push quick coupler from
“IE OUT”
tube onto the cartridge
“OUT”
coupler fitting, until it clicks. (See
Figure 2)
(4) Push quick coupler from
“IE IN” tube
onto the cartridge
“IN” coupler fitting,
until it clicks. (See Figure 2)
●
Removal procedure:
(1) Remove quick coupler by pushing
the black part of the coupler casing
toward the cartridge while pulling the
coupler off. (See Figure 3)
(2) Remove the coupler from the IN side
first, followed by the OUT side.
Figure 1
Figure 2
Figure 3
※
Quick couplers may be stiff with first few uses and cartridge coupler fittings may break off, if
too much force is applied. If installing the couplers is too difficult with ion exchange cartridge
installed, remove it from the main unit and connect the hoses, first. Be sure not to kink the
plastic tubes while using this method.
※
Once hoses are connected, give each coupler a gentle tug to verify that they are securely in
place.
※
When removing quick couplers, water may drip from the ion exchange resin cartridge fitting.
Be careful not to let water get on the leak sensor.
※
Be sure not to cross the hoses (e.g. connect IN hose to OUT tip and vice versa) when
replacing the ion exchange cartridge. Malfunction may result.
Slide
Содержание Pure Line WE200
Страница 50: ...48 11 WIRING DIAGRAM ...