Net Weighing - Tare Operation

Net weighing displays the weight of an item without the weight of its container.  The weight of the con-

tainer is tared off.

 

 Caution

 

 

The total gross weight should never exceed the capacity of the scale.  This can damage the load 

cell.

1.  Begin with nothing on the weighing platform, the display reading zero, 

and the center of zero and stable indicators lit.

2.  Place an empty container on the weighing platform and wait for the 

stable indication.

3.  Press the 

 key.  The weight of the container will be subtracted au-

tomatically and the net weight indicator will display.

4.  Place the items to be weighed in the container.  The weight displayed 

will be the net weight of the items.

5.  To display the gross weight, press the 

 key.  The net weight indica-

tor will no longer display, and the displayed weight will be the gross 
weight of the container and the items.  Press the 

 key again to 

return to the net weight.

6.  To return to gross weighing mode, remove all items from the weighing 

platform, wait for the stable indication, and press the 

 key.

IV Using the Scale

9

Содержание PPC-300

Страница 1: ...TO CORPORATION 1775 S Murray Blvd Colorado Springs CO 80916 USA Tel 719 591 1500 Fax 719 591 1045 YAMATO TECH CORPORATION 112 19425 Langley By Pass Surrey B C V3S 6K1 Canada Tel 604 533 2338 Fax 604 533 0827 www YamatoCorp com NTEP CC 08 041 Version 2 01 ...

Страница 2: ......

Страница 3: ...ng the Batteries 7 Installing the AC Adaptor 7 Leveling the Scale 7 IV Using the Scale 8 Turning the Scale On and Off 8 Zero the Scale 8 Selecting Units 8 Checking the Battery Level 8 Setup Mode Indicator 8 Basic Weighing 8 Net Weighing Tare Operation 9 Batch Weighing 10 Selectable Functions 11 Packing Function 12 Checkweighing Function 15 Grading Function 18 V User Parameters 21 VI Troubleshootin...

Страница 4: ...ng Warns users about a procedure that could result in injury or property damage if not performed properly or if ignored Caution Warns users about a procedure that could result in minor injury or damage to the scale if not performed properly or if ignored Indicates an action that must never be performed Indicates an action that must always be performed Note Statements that provide additional inform...

Страница 5: ...f the scale Do not twist the platform This can damage the scale Do not shock load the scale This can damage the scale Do not push the indicator or keys with sharp objects They can puncture or break the switch membrane panel This scale is a sensitive weighing instrument avoid physical shocks If you drop something on the scale overload the scale step on the platform or drop the scale the scale may b...

Страница 6: ... large motors This will affect the accuracy The operating temperature range is from 10 C to 40 C 14 F to 104 F Do not subject the scale to sudden temperature changes Allow the scale to adjust to new temperatures before use If the scale is sealed do not break the seal If you break the seal the scale will not be considered legal for trade In this case contact your dealer Do not disassemble or modify...

Страница 7: ...ss and Tare keys simultaneously to toggle through the available weight units 7 User Parameter Mode Indicates that the scale is in user parameter mode 8 Test Mode In normal weighing mode indicates that the scale is not configured to be sealed for legal for trade operation In test mode indicates that the scale is in test mode 9 Battery Status Indicates approximate remaining battery capacity in third...

Страница 8: ...g x 0 01 kg 60 lb x 0 02 lb 1056 oz x 0 5 oz 66 lb x 0 5 oz Capacity Combined units such as pound ounce are not legal for trade Weight display PPC 300 single display PPC 300D dual display Type 7 segment LCD Character size etc 0 5 13 5 mm W x 1 2 30 mm H 5 digits Functions One touch tare battery charge indicator auto off timer and check weighing packing and grading functions Housing ABS resin Optio...

Страница 9: ...YCO 1125 03 13 PPC 300 PPC 300D Operation Manual 6 II Description Specifications Dimensions 228 200 242 198 178 288 113 120 ...

Страница 10: ...ling the AC Adaptor Use only the AC adaptor specified for use with this scale Using the wrong adaptor can damage the scale Gently and carefully turn the scale on its side Open the battery cover and remove any bat terres that are installed The scale will continue to draw power from the batteries if they are installed even with an AC adapter connected Securely close the battery cover Remove the seal...

Страница 11: ...and pound ounce units press the and keys at the same time Checking the Battery Level The battery status indicator shows the approximate life remaining in the batteries as follows 2 3 to full charge remaining 1 3 to 2 3 charge remaining Less than 1 3 charge remaining Battery charge is too low Replace all four D batteries observing the polarity markings Setup Mode Indicator If the setup mode indicat...

Страница 12: ...n the weighing platform and wait for the stable indication 3 Press the key The weight of the container will be subtracted au tomatically and the net weight indicator will display 4 Place the items to be weighed in the container The weight displayed will be the net weight of the items 5 To display the gross weight press the key The net weight indica tor will no longer display and the displayed weig...

Страница 13: ...y The weight of the container will be subtracted au tomatically the net weight indicator will display and the scale will read zero 4 Add the first item to be weighed in the container until the desired weight is reached 5 Press the key to return to net zero Add the second item to be weighed in the container until the desired weight is reached 6 Press the key to return to net zero Continue adding it...

Страница 14: ... not legal for trade Parameter 01 002 The checkweighing function uses a lower limit and upper limit to establish an acceptable weight range for portioning quality control and filling while displaying the weight loaded on the platform The bar chart gives a graphical representation of the status of the loaded weight in relationship to the upper and lower limits for quick reference Up to 99 different...

Страница 15: ...ly the displayed value 3 To store any changes made to this parameter and to advance to the next parameter press the key The right arrow next to this key indicates that it advances the display to the next parameter 4 If you wish to set the automatic tare advance to parameter 06 000 disables the automatic tare 001 enables the automatic tare Press the key to save any changes and advance to the next p...

Страница 16: ...to select a digit to change and the and keys to decrease or increase the digit s value Example Press the key three times to select the fourth digit Press the key once to change the digit to 1 5 Press the key to enter the target weight and progress to the ac ceptable weight range The acceptable weight range is the value above the target weight that is acceptable Example For an accept able range of ...

Страница 17: ...with the product to be weighed As the amount of product increases the negative weight reading will ap proach zero and the length of the bar chart will increase 4 The weight reading will begin to flash and the bar chart will enter the ACCEPT range when the target weight is reached 5 The weight reading will continue to flash and the bar chart will remain in the ACCEPT range through the acceptable we...

Страница 18: ...e the parameter value to 002 The down and up arrows next to these keys indicate that they decrease and increase respectively the displayed value 3 To store any changes made to this parameter and to advance to the next parameter press the key The right arrow next to this key indicates that it advances the display to the next parameter 4 If you wish to set a flashing weight reading advance to parame...

Страница 19: ...4 Use the key to select a digit to change and the and keys to decrease or increase the digit s value Example Press the key three times to select the fourth digit Press the key once to change the digit to 1 5 Press the key to enter the lower limit weight and progress to the upper limit weight The acceptable weight range is any weight be tween the upper and lower limits inclusive 6 Use the key to se...

Страница 20: ...g the scale with the product to be weighed As the amount of product increases so will the weight display and the length of the bar chart 4 The weight reading will begin to flash and the bar chart will enter the ACCEPT range when the lower limit weight is reached 5 The weight reading will continue to flash and the bar chart will remain in the ACCEPT range up to and including the upper limit weight ...

Страница 21: ...ncrease respectively the displayed value 3 To store any changes made to this parameter and to advance to the next parameter press the key The right arrow next to this key indicates that it advances the display to the next parameter 4 The second parameter determines the number of grades for all setups The greater the number of grades the fewer setups that can be stored in the scale s memory Setting...

Страница 22: ...etting If the configura tion is already setup then press the key to exit setup and begin weighing The single bar on the bar chart indicates that this is the first grade weight 4 Use the key to select a digit to change and the and keys to decrease or increase the digit s value Example Press the key eight times to change the digit to 8 Press the key to save this weight and advance to the next grade ...

Страница 23: ...e displayed and the center of zero indicator should not 3 Load the scale with the product to be graded If the product falls at or above the first grade setpoint and below the second grade setpoint then grade 01 will be displayed 4 If the product falls at or above the second grade setpoint and below the third grade setpoint then grade 02 will be displayed 5 If the product falls at or above the thir...

Страница 24: ...Disabled 001 Enabled packing function only 07 000 Flashing display 000 Disabled 001 Flash when under weight 002 Flash for acceptable weight 003 Flash when over weight 004 Flash when under or over weight 005 Flash when buzzer sounds 08 19 000 Inactive 000 Do not adjust these settings 22 000 Vibration compensation 000 Standard highest response speed 001 Increased slower response speed 002 Maximum sl...

Страница 25: ...e will display two dashes in place of the grade value until the weight reading is stable Then it will display the appropriate grade 05 Auto off Timer The PPC 300 series of scales can be set to automatically turn off after a specified time with no activity This feature only operates when the scale is using batteries for power 06 Automatic Tare The packing function can be set to automatically tare a...

Страница 26: ...tteries Scale will not turn on Dead batteries or AC DC adapter not connected Replace all four batter ies with fresh batteries or verify proper adapter is connected to the scale and plugged into a pow ered outlet Load over allowable limit at start up Remove all items from the scale platform and press the key Load under allowable limit at start up Remove items wedged under the scale platform and pre...

Страница 27: ...pan on the scale platform if removed Remove all items from the scale platform and press the key Sensor error Turn the scale off wait one minute and then turn the scale back on Circuit error Turn the scale off wait one minute and then turn the scale back on Controller error Turn the scale off wait one minute and then turn the scale back on Additional troubleshooting steps require the services of a ...

Страница 28: ...CORPORATION 1775 S Murray Blvd Colorado Springs CO 80916 USA Tel 719 591 1500 Fax 719 591 1045 YAMATO TECH CORPORATION 112 19425 Langley By Pass Surrey B C V3S 6K1 Canada Tel 604 533 2338 Fax 604 533 0827 www YamatoCorp com ...