4
2. Before operating the unit
Precautions for Installation
Warning
1. Be sure to connect the ground wire.
・
To prevent electric shock accidents due to fault current, be sure to connect the ground
wire to the ground terminal on the building side. This machine falls in the category of
Class D Ground Work.
・
Never connect the ground wire to a gas pipe, water pipe, ground wire of a telephone or
lightning conductor. Hazard of a fire and electric shock.
2. Select a suitable installation site.
Do not install the machine in the following places:
・
An unstable place.
・
A place where flammable or corrosive gas is produced.
・
A place where ambient temperature rises above 35
℃
.
・
A place where temperature changes largely.
・
A place filled with dust or a humid place.
・
A place under direct sunlight.
・
A place that is subjected to vibrations.
3. Install the machine on a level foundation.
Install the machine on a level foundation. If the machine is not in contact with the floor
uniformly, vibration and noise may be produced. Also unexpected troubles or failures may
occur.
4. Use a suitable distribution panel.
Use a distribution panel that is suitable for the electric rating of the machine.
Electric rating: 3-phase, 200 VAC (Differ depending on the region. See page 19.), 15 A,
50/60 Hz
If the capacity is insufficient or the machine is used on a line to which other equipment is
connected, the machine may not function properly or the circuit breaker in the distribution
panel may operate.
Содержание PDC610
Страница 40: ...37 13 WIRING DIAGRAM...
Страница 41: ...38 13 WIRING DIAGRAM...