25
Operation Method
Input Program
1
Select program function
Check that the power in turned on.
PROGRAM
TROUBLE
ENTER
CANCEL
Select step
S01
Program OPR
①
Press the PROGRAM key. The program menu
starts and step number selection screen is
displayed.
Press the "
▼▲
". The registered step numbers
and the smallest number of un-used step are
displayed in sequence. Select the step number
among them.
The "S01" is displayed when no program is
registered. In this case, the "
▼▲
" are invalid.
②
Press the ENTER key. The selected step
number is decided and the setting item selection
screen is displayed. The "Set TEMP" is
displayed first.
③
Press the CANCEL key to cancel the program
menu.
2
Edit step
The following four setting items are included in a step.
The unit operates with fixed rate inclination if the "START TEMP" and the "Set Temp" are different.
It enters into "Wait" if the temperature does not reach to the setting temperature within a setting
time.
Set Temp.
TEMP
.
TIME
Set Time
TEMP
.
START
.
(Operation with fixed rate inclination)
Setting item
Setting range
Setting temperature
Setting temperature range by product type
Setting time
0 minute to 999 hour and 59 minutes, End
Repeat initiating step
None, Registered step numbers 1 to 32
Repeat count
1 to 9999, Infinity