17
Operation Method
Auto Stop Operation
As shown in the following figure, the device stops operating automatically by setting the timer.
Setting of the Auto Stop Operation
1
Turn on the power
POWER
Turn on the power switch of the unit (earth leakage
breaker). Pressing the POWER key turns on the
power. The Main Display indicates the
temperature in furnace. The Sub Display
indicates "Standby" and the STANDBY lamp lights
on.
2
Select operation mode
MENU
①
Pressing the MENU key displays the operation
mode selection screen.
②
The Sub Display on the operation mode selection
screen displays the name of operation mode
currently
selected
with
blinking.
The
corresponding operation mode lamp blinks at the
same time.
③
Keep pressing the MENU key until the auto stop
operation mode is displayed.
④
Press the ENTER key. The auto stop operation
mode is decided.
The fixed temperature operation is selected at the
initial setting of unit. The operation mode carried
out last is selected in case other than it.
Operation
period
SV:
Setting temperature
t:
Time
t
Set timer
▼
SV
Activate timer
Start operation (manual)
Stop operation (auto)
Waiting