YCO-201008
DP-6700
Technical Manual
+10 operation for parameter value:
Press the
key while holding down the
key to
increment the parameter value by 10.
If the setting range of the parameter is less than 10, "0000"
stays on the panel in spite of the operation.
-10 operation for parameter value:
Press the
key while holding down the
key to dec-
rement the parameter value by 10 (or increment it by -10).
If the settings range of the parameter is less than 10, the
maximum setup limit for the parameter stays on the panel
in spite of the operation.
(2) System Parameter Mode (to set stability detection and others)
Related page: Page
1-15 1.3.1 How to enter test mode and meanings of panel indications
System parameter mode is used to change dealer parameters such as stability detection. The following
parameters from 30 to 39 and B1 to C3 can be set, confirmed and changed.
In this mode, (1) user parameters fromt 01 to 29 can be set, confirmed and changed.
To switch to dealer parameter mode, you must switch to test mode first. See
1.3.1 How to enter test
mode and meanings of panel indications
on Page
1-5
for switching to test mode.
When you replace the CPU board, make sure to check if the parameters settings are correct, referring
to the parameter list.
System Parameters
No.
Item
Value
Function
30
System ID
016:
Prohibit to Change
31
Stable - State Sampling Count
000-015:
*Use the time of "setting value X 2"
32
Stable - State Count
000-050:
33
Very Stable - State Count
000-050:
34
Stable - State Releasing Count
000-050:
35
Stable - State Average Count
000-015:
37
Data Accumluation at Multi-
Function
000:
001:
No Data Accumulation
Data Accumulation
38
Local Multi-Function setting dur-
ing communication
000:
001:
Editable
No Editable
39
Skip in Reading on Display
000:
001-015:
Skip in Reading Disable
Skip in Reading at specified
1-11
Chapter 1 Adjustment method
Содержание DP-6700
Страница 2: ...YCO 201008 DP 6700 Technical Manual ...
Страница 28: ...YCO 201008 DP 6700 Technical Manual 1 25 Chapter 1 Adjustment method ...
Страница 29: ...YCO 201008 DP 6700 Technical Manual 1 26 Chapter 1 Adjustment method ...
Страница 30: ...YCO 201008 DP 6700 Technical Manual 1 4 Wiring diagram 1 27 Chapter 1 Adjustment method ...
Страница 32: ...YCO 201008 DP 6700 Technical Manual 1 29 Chapter 1 Adjustment method ...
Страница 33: ...YCO 201008 DP 6700 Technical Manual 1 30 Chapter 1 Adjustment method ...
Страница 38: ...YCO 201008 DP 6700 Technical Manual Chapter 3 How to disassemble 3 1 ...
Страница 39: ...YCO 201008 DP 6700 Technical Manual Chapter 3 How to disassemble 3 2 ...
Страница 40: ...YCO 201008 DP 6700 Technical Manual Chapter 3 How to disassemble 3 3 ...
Страница 41: ...YCO 201008 DP 6700 Technical Manual Chapter 3 How to disassemble 3 4 ...
Страница 54: ...YCO 201008 DP 6700 Technical Manual ...