24
In the Event of Failure…
Trouble Shooting
Condition
Possible Causes
Refrigerator does not start when
turning on the power switch.
Power plug is not connected to the receptacle correctly.
Power failure.
Earth leakage breaker is turned to
“OFF”
Not fallen the temperature.
The setting is high than the in-bath temperature.
The condenser filter at the lower front of the main body is dirty.
The condenser fins are clogged.
Heat load of circulating liquid is too large.
The environmental temperature is high.
The Suction hole is blocked.
Refrigerator cannot be restarted.
The refrigerator is in overloaded. Turn off the power of the
refrigerator immediately, keep the temperature, and turn on
the power of the refrigerator after a while.
The liquid is not circulated.
The circulation path is closed or throttled too narrow at any
point.
The argillaceous or specific gravity of the cooling liquid
appropriate.
Error Display
Error Sign
Cause
Remedy
UUUU
・
Sensor disconnection
・
Coolant temperature exceeded
35 degrees Celsius
・
Check the sensor connection.
・
Please do below coolant temperature 35
degrees Celsius
FALR
・
Malfunction of the temperature
controller
・
Stop operation.
Turn off the power immediately,
In the case if the error other than listed above occurred, turn off the power switch and primary power
source immediately. Contact the shop of your purchase or nearest Yamato Scientific Service Office.