Indexes
9.1
Explanation of the messages
This chapter contains a list of all the message codes and related
message texts for the SensoLyt
®
700 IQ H sensor.
All message codes of the SensoLyt
®
700 IQ H end with the number
"311".
9.1.1 Error messages
Information on
the contents and structure of the log book and
the structure of the message code
See IQ
S
ENSOR
N
ET
system operating manual, chapter
L
OG
BOOK
.
Message code
Message text
EA1311
Meas. range exceeded or undercut
* Check process
* Select other meas. range
EA2311
Sensor temperature too high!
* Check process and application
EA3311
Sensor temperature too low!
* Check process and application
EC1311
Sensor could not be calibrated,
Sensor blocked for measurement
* Check calibration conditions and calibration standard
* View calibration history
* Service sensor immediately
(see operating manual)
EI3311
Operational voltage too low
* Check installation and cable lengths, Follow installation instructions
* Power supply module overloaded
* Check terminal and module connections
* Defective component, replace components