4.2.5 Calibration result
Calibration evaluation
After calibrating, the calibration data and current state of the sensor are
evaluated automatically. The asymmetry and slope are evaluated sep-
arately. The values must be within the following ranges:
Slope:
-50 ... -62 mV/pH
Asymmetry:
-45 ... +45 mV
If one of the two values is outside the specified range the calibration is
evaluated as not successful, i.e. the sensor could not be calibrated.
A calibration procedure can have the following results:
Possible calibration
results
Successfully calibrated.
End of the CAL_CON_1P calibra-
tion.
The values determined for
Slope
and
Asymmetry potential
are dis-
played.
Calibration is finished.
Confirm with
<OK>
.
The display returns to the mea-
sured value display.
Display
Explanation
Display after
calibrating
Log book entries
(meaning/actions)
Measured value
display
Sensor was successfully calibrated.
Calibration data see calibration history.
"----"
Sensor could not be calibrated.
Sensor blocked for measurement.
– Carry out maintenance activities immedi-
ately (see operating manual).
– View the calibration history.
– Check the calibration conditions and calibra-
tion standard.
Information on the contents and structure of the log book,
and how to call it up, is given in the Log book chapter of the
IQ
S
ENSOR
N
ET
system operating manual.