background image

40

41

Part Number

Description

605202

Galvanic DO Sensor

605203

Polarographic Sensor

603059

Flow cell, 203 ml 

603056

Flow cell mounting spike

603075

Carrying case, large, soft-sided

603162

Carrying case, small, soft-sided (holds instrument 
and up to a 4 meter cable/probe assembly

603074

Carrying case, hard-sided

603069

Belt clip for clipping instrument onto belt

063517

Ultra clamp for instrument for clamping 
instrument to lab counter or other surface

063507

Tripod for instrument

603062

Cable management kit, included with all cables 
longer than 1 meter

605978

Cable weight, 4.9 oz, stackable

603070

Shoulder strap

003821

pH 4 Buffer, box of 6 pints

003822

pH 7 Buffer, box of 6 pints

003823

pH 10 Buffer, box of 6 pints

603824

pH Buffer, assorted case, 2 pints each of buffer 
4, 7 and 10

605306

5908 Cap membrane kit, 1.25 mil PE (6 each), 
includes electrolyte (yellow cap). Used with 
Polarographic sensor. 

605307

5909 Cap membrane kit, 2 mil PE (blue cap, 6 
each), includes electrolyte (blue cap). Used with 
Polarographic sensor. 

605913

5913 Cap membrane kit, 1.25 mil PE (yellow cap, 
6 each), includes electrolyte. Used with Galvanic 
sensor. 

605914

5914 Cap membrane kit, 2 mil PE (blue cap, 6 
each), includes electrolyte. Used with Galvanic 
sensor. 

*All cables include a temperature sensor. The pH or ORP and Dissolved 
Oxygen sensors are sold separately.  

D E C L A R A T I O N   O F   C O N F O R M I T Y

The undersigned hereby declares on behalf of the named manufacturer 
under our sole responsibility that the listed product conforms to the 
requirements for the listed European Council Directive(s) and carries the 
CE mark accordingly.

Manufacturer:

YSI Incorporated
1725 Brannum Lane
Yellow Springs, OH 45387
USA

Product Name:

Pro1020 Water Quality Instrument

Model Numbers
Instrument/
Accessory:

Pro1020 (6051020) 

Probe/Cable 
Assemblies:

6051020-1, -4, -10, -20, and -30

Conforms to the following:
Directives:

EMC 2004/108/EC
RoHS 2011/65/EU
WEEE 2012/19/EU

Harmonized 
Standards:

• 

EN61326-1:2006 (IEC 61326-1:2005)

• 

IEC 61000-3-2:2005

• 

IEC 61000-3-3:2005

Supplementary 
Information:

All performance met the operation 
criteria as follows:
1. ESD, IEC 61000-4-2:2001
2. Radiated Immunity, IEC 61000-4-3:2006
3. Electrical Fast Transient (EFT), IEC 61000-
4-4:2004, +Corr. 1:2006 + Corr. 2:2007
4. Radio Frequency, Continuous Conducted 
Immunity, IEC61000-4-6:2006
5. IEC 6100-4-8:2001

Authorized EU 
Representative

Xylem Analytics UK Ltd
Unit 2 Focal Point, Lacerta Court, Works Road
Letchworth, Hertfordshire, SG6 1FJ UK

Signed:  Lisa M. Abel          

 

             Date:  23 January 2014

Title:  Director of Quality
 

Содержание YSI Pro1020

Страница 1: ...USER MANUAL English Fran ais Espa ol Portugu s Pro1020 User Manual DOCUMENT 605187...

Страница 2: ...1 Battery Installation 1 Key Pad 2 Connecting the Sensor and Cable 3 Membrane Installation 6 Run Screen 7 Backlight 7 Powering Off 7 Navigation 7 First Power On 8 System Setup Menu 9 DO Local 10 Last...

Страница 3: ...YSI will repair or replace at its sole discretion free of charge any product that YSI determines to be covered by this warranty To exercise this warranty call your local YSI representative or contact...

Страница 4: ...arts or materials are damaged or missing contact YSI Customer Service at 800 897 4151 1 937 767 7241 or the authorized YSI distributor from whom the instrument was purchased BATTERY INSTALLATION The i...

Страница 5: ...he bottom of the Run screen and to increase numerical inputs 3 Power and Backlight Press once to turn instrument on Press a second time to turn backlight on Press a third time to turn backlight off Pr...

Страница 6: ...Next remove the o ring and slide the lid off the sensor INSTALLING THE DISSOLVED OXYGEN SENSOR Before installing a sensor or connecting the cable to the instrument the Sensor Type must be configured f...

Страница 7: ...cap from the dissolved oxygen sensor by pulling it straight off Thoroughly rinse the sensor tip with distilled or deionized water Fill the cap membrane 3 4 full of electrolyte solution then tap the ca...

Страница 8: ...dissolved oxygen sensor type and membrane type Use the up or down arrow keys to highlight the appropriate language sensor and membrane then press enter to confirm figure 6 The Sensor Type must be con...

Страница 9: ...instead of during calibration Example Instrument 1 with DO Local enabled At 737 mmHg barometric pressure roughly 841 ft above sea level the instrument would calibrate to 100 When taking measurements...

Страница 10: ...ompatible sensors for use with a field cable Polarographic This sensor has a black sensor body and is engraved with the model number 2003 Polarographic will be abbreviated Polaro in the instrument men...

Страница 11: ...lay steadily and the instrument will beep twice if Audio is turned on DO Auto Stable can be set to a change of 0 0 to 1 9 over 3 to 19 seconds ISE Auto Stable can be set to a change of 0 0 to 9 9 over...

Страница 12: ...nge the language Highlight the desired language English Spanish Portuguese or French and press enter to enable The enabled language will have an X in the box next to it Highlight the ESC Exit box and...

Страница 13: ...ate formula for this uncorrection is True BP Corrected BP 2 5 Local Altitude in ft above sea level 100 Although the Pro1020 barometer range is 400 0 to 999 9 mmHg you can not adjust the value across t...

Страница 14: ...ubleshooting guide for possible solutions CALIBRATING IN PERCENT DO LOCAL ENABLED Perform this calibration procedure when DO Local is enabled and Quick DO Cal is disabled in the System Setup menu 1 Pe...

Страница 15: ...5 If necessary use the up and down arrow keys to adjust the pH buffer value Note the pH mV reading which ideally should be between 50 and 50 in buffer 7 6 Press enter to continue to second point 7 Ri...

Страница 16: ...membrane If using the 1001A amplified pH sensor be sure to use the guard extender supplied in the 605216 amplified pH sensor kit The extender is screwed onto the sensor bulkhead before the guard or c...

Страница 17: ...enter to regain access to scrolling through the saved data The data set displayed is indicated by the data set number 01 through 50 ERASING DATA While viewing saved data press the enter key to access...

Страница 18: ...or connector Check the o ring and the groove for any excess grease or contamination If contamination is evident clean the o ring and nearby plastic parts with lens cleaning tissue or equivalent lint f...

Страница 19: ...the sensor in clean water wipe with a cotton swab saturated with clean water not DI water and then rerinse with clean water To be certain that all traces of the acid are removed i from the sensor cre...

Страница 20: ...d in the membrane kit Do not sand the electrodes every membrane change as this is not routine maintenance In fact visually the anode may appear tarnished and operate properly YSI recommends using the...

Страница 21: ...embrane should be changed if a bubbles are visible under the membrane b dried electrolyte is visible around the membrane and c if the sensor shows unstable readings or other sensor related symptoms Th...

Страница 22: ...em for service Symptom Possible Solution Instrument will not calibrate pH or ORP instrument displays Calibration Over Calibration Under or Unstable Reading during calibration 1 Verify correct ISE sens...

Страница 23: ...specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website at ysi com or contact YSI Tech Suppor...

Страница 24: ...E blue cap 6 each includes electrolyte Used with Galvanic sensor All cables include a temperature sensor The pH or ORP and Dissolved Oxygen sensors are sold separately D E C L A R AT I O N O F C O N F...

Страница 25: ...country and region and users are expected to understand and follow the battery disposal requirements for their specific locale C O N TA C T I N F O R M AT I O N ORDERING AND TECHNICAL SUPPORT Telephon...

Страница 26: ...Item 605187 Rev A January 2014 2014 YSI Incorporated...

Отзывы: