TruLab pH/ISE 1320
Settings
ba76161e01
08/2015
73
9.1.4 Settings for ORP measurements
The settings for ORP measurements are made in the menu for calibration and
measurement settings of the pH/ORP measurement. To open the settings, dis-
play the required measured parameter in the measured value display and
press the
<MENU/ENTER>
key. After completing the settings, switch to the
measured value display with
<M>
.
In the following table, only those settings are listed that influence the ORP mea-
surement.
Default settings are printed in
bold
.
Menu item
Possible
setting
Description
Alternative temper-
ature
On
Off
Takes the temperature value from
the 2nd sensor.
Man. temperature
-25 ...
+25
...
+130 °C
Entry of the manually determined
temperature. For measurements
without temperature sensor only.
Resolution mV
0.1
1
Resolution of the mV display
Reset
-
Resets all sensor settings to the
delivery condition (see section 9.4.1
R