4 - 2
ba76178e01
10/2014
Measuring / Operation
FDO
®
70x IQ H
With air temperatures under 5 °C we recommend performing the
function check and user calibration not in air but in air-saturated
water that has a higher temperature. You obtain air-saturated water
by pouring water several times in and out of two vessels so that it
sparkles.
4.2.2
Function check
A function check is the simplest way to determine whether the sensor
needs to be cleaned or user-calibrated.
Principle
The function check can either be done in water vapor-saturated
air or in air-saturated water (see C
on
Steps
Generally, a check on the IQ
S
ENSOR
N
ET
is carried out as follows. Sys-
tem specific details are given in the respective IQ
S
ENSOR
N
ET
system
operating manual.
2 cm
1
Switch to the measured value display with
<M>
and select the
FDO
®
70x IQ H sensor.
2
Press
<C>
.
The next step switches on the maintenance condition for the
sensor. A corresponding message appears on the display.
3
Confirm the note with
<OK>
.
The maintenance condition is active.
4
Select the
TEST
procedure and press
<OK>
.
5
Put the sensor into the calibration position (water vapor-satu-
rated air or air-saturated water - see section 4.2.1).
6
Press
<OK>
.
The sensor starts the check. The display switches to the mea-
sured value display. The
CAL
indicator flashes instead of the
main measured value. At the same time, the momentary rela-
tive slope flashes as the secondary measured value with the
addition of
TEST
. The process ends automatically as soon as
the measured values meet the criterion for the stability control.
With a great temperature difference between the sensor and
the environment this may take a while. Subsequently, the main
measured value and temperature are displayed.
Содержание YSI FDO 700 IQ H
Страница 8: ...1 4 ba76178e01 10 2014 Overview FDO 70x IQ H...
Страница 12: ...2 4 ba76178e01 10 2014 Safety FDO 70x IQ H...
Страница 18: ...3 6 ba76178e01 10 2014 Commissioning FDO 70x IQ H...
Страница 26: ...4 8 ba76178e01 10 2014 Measuring Operation FDO 70x IQ H...
Страница 32: ...5 6 ba76178e01 10 2014 Maintenance cleaning replacement parts FDO 70x IQ H...
Страница 40: ...7 6 ba76178e01 10 2014 Technical data FDO 70x IQ H...
Страница 42: ...8 2 ba76178e01 10 2014 Contact Information FDO 70x IQ H...
Страница 47: ......