Lists
Alyza IQ NH4
162
ba76202d02
05/2020
8
Lists
8.1
Explanation of the messages
This chapter contains a list of all message codes and the related message texts
that can occur in the log book of the IQ S
ENSOR
N
ET
system for the Alyza IQ an-
alyzer.
The last three digits of the message code form the component code. It identifies
the component (active component) that caused the message:
Some error messages contain an internal error code, starting with "#".
8.1.1 Error messages
Information on the contents and structure of the log book and the
structure of the message code is given in the L
OG
BOOK
chapter of
the IQ S
ENSOR
N
ET
system operating manual.
Module code
Component
3E1
Alyza IQ NH4, channel 1
3E2
Alyza IQ NH4, channel 2
552
MIQ/Alyza (adapter ADA)
Message code
Message text
EAM3Ex
Meas. range exceeded or undercut
* Check process
EC1552
Automatic calibration error: Calibration standard could not be deter-
mined or is not suitable for current measurement range. Alyza IQ
stopped!
* check selected measuring range
* check calibration standard used
ES13Ex
Component hardware defective
* Contact service
ES2552
Component hardware defective xxx
* Contact service
ES3552
Danger due to condensation water forming in the measuring unit.
The measuring unit is switched off.
* Dry the measuring unit
ES4552
Error temperature control of the measuring unit.
* Contact Service