4.5.5 Standard adjustment (user calibration)
With some of the programs (methods) for concentration measurement, you
can optimize the calibration curve stored in the meter with the aid of the
Stan-
dard adjustment
function.
A standard adjustment is only valid if the deviation compared to the original
calibration is no more than 30%.
A standard adjustment is always stored for the program that is presently
called up. A standard adjustment is only deleted if
a new standard adjustment is carried out
the standard adjustment is erased manually
the meter is reset to delivery status (
Reset
function)
Standard adjust-
ment
8 Confirm the result with
<START/ENTER>
.
The blank value measurement is completed.
The meter is ready to measure.
or:
Discard the result with
<ESC>
.
Subsequently, carry out a new blank value measurement.
i
83: A6/25 MC
Cell
i
= 16 mm
i
Absorbance
= 0.301
i
Blank value mea-
surement
successful!
Measure blank value
1 If necessary, press the
<M>
key several times until the measured pa-
rameter
Concentration
is selected.
2 Select a program with
<PROG>
as necessary.
The following measurement of the standard applies only to the
selected program.
3 Open the
Photometry \ Adjustment
menu with
<CAL/ZERO>
.