Settings
pH/Cond 3320
48
ba77054d01
02/2014
Setting the calibra-
tion interval
The calibration interval is set to 7 days in the factory.
You can change the interval (1 ... 999 days):
8.1.4
Settings for ORP measurements
The settings for ORP measurements are made in the menu for calibration and
measurement settings of the pH/ORP measurement. To open the settings, dis-
play the required measured parameter in the measured value display and
press the
<ENTER>
key. After completing the settings, switch to the measured
value display with
<M>
.
In the following table, only those settings are listed that influence the ORP mea-
surement.
Default settings are printed in
bold
.
To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.
1.
Using
<ENTER>
, open the menu for the measurement settings.
2.
In the
Calibration
/
Calibration interval
menu, set the calibration interval
with
<
><
>
.
3.
Confirm the setting with
<ENTER>
.
4.
Quit the menu with
<M>
.
Menu item
Possible
setting
Description
Man. temperature
-25 ...
+25
...
+130 °C
Entry of the manually determined
temperature. For measurements
without temperature sensor only.
Resolution mV
0.1
1
Resolution of the mV display
Reset
-
Resets all sensor settings to the
delivery condition (see section 8.3.1
R