Multi 3630/3620 IDS
Firmware update for OxiTop
®
-IDS measuring heads
ba77045e01
09/2019
33
8
Firmware update for OxiTop
®
-IDS measuring
heads
Available firmware updates for OxiTop
®
-IDS measuring heads are provided on
the Internet. With the "Firmware Update " program and a PC you can update the
firmware of a OxiTop
®
-IDS
measuring head to the newest version.
For the update you have to connect the Multi 3630/3620 IDS meter to a PC via
the USB interface and start the update program at the PC.
The update requires:
a free USB interface (virtual COM port) on the PC
the current driver for the USB interface
(see operating manual of your meter)
the current firmware for the meter
(see operating manual of your meter)
the current firmware for the WLM-M adapter
(see operating manual of your WLM system)
the USB cable (included in the scope of delivery of the Multi 3630/3620 IDS).
After switching the Multi 3630/3620 IDS meter off and on you can check whether
the OxiTop
®
-IDS measuring head has taken over the new software version (see
Multi 3630/3620 IDS operating manual).
1
Install the downloaded firmware update on a PC.
An update folder is created in the Windows start menu.
If an update folder for the OxiTop
®
-IDS measuring head is already exist-
ing, the new data are displayed there.
2
In the windows start menu, open the update folder and start the firm-
ware update program for the OxiTop
®
-IDS measuring head.
3
Using the USB interface cable, connect the Multi 3630/3620 IDS to a
USB interface (virtual COM port) of the PC.
4
Switch on the Multi 3630/3620 IDS.
5
At the OxiTop
®
-IDS measuring head in the
Service
menu, start the
update (with
Update mode
, see operating manual OxiTop
®
-IDS mea-
suring head). The firmware is loaded and automatically installed.
6
In the firmware update program at the PC, start the update process with
OK.
7
Follow the instructions of the firmware update program.
During the programming process, a corresponding message and a
progress bar (in %) are displayed.
The programming process takes up to 5 minutes. A terminatory mes-
sage is displayed after a successful programming process. The firm-
ware update is completed.
8
Disconnect the Multi 3630/3620 IDS from the PC.
The meter and OxiTop
®
-IDS measuring head are ready for operation
again.