Dissolved oxygen
Multi 3630 IDS
48
ba77170e03
10/2017
7
Dissolved oxygen
7.1
Measuring
7.1.1 Measuring D.O.
Selecting the
displayed
measured parameter
You can switch between the following displays with
<M>
:
D.O. concentration [mg/l]
D.O. saturation [%]
D.O. partial pressure [mbar].
Salinity correction
When measuring the concentration of solutions with a salt content of more than
1 g/l, a salinity correction is required.
You have the following options to measure the salinity:
Measurement of the salinity with an IDS conductivity sensor.
If the measured value from an IDS conductivity sensor is taken over, the sta-
tus indicator [Sal
↑
] is displayed in the measurement window of the IDS oxy-
gen sensor. The status indicator [Sal
↓
] is displayed in the measurement
window of the IDS conductivity sensor.
Manual determination and input of the salinity.
The sensor connection and the USB-B (device) interface are gal-
vanically insulated. This facilitates interference-free measurements
also in the following cases:
Measurement in grounded test samples
Measurement with several sensors connected to one
Multi 3630 IDS in one test sample
1
Connect the D.O. sensor to the meter.
The D.O. measuring screen is displayed.
2
Check or calibrate the meter with the sensor.
3
Immerse the D.O. sensor in the test sample.
30.09.2015 08:00
Info
Содержание WTW OxiTop-IDS Multi 3630 IDS
Страница 1: ...OPERATING MANUAL ba77170e03 10 2017 Multi 3630 IDS DIGITAL METER FOR IDS SENSORS 3630 IDS...
Страница 2: ...Multi 3630 IDS 2 ba77170e03 10 2017 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Страница 15: ...Multi 3630 IDS Commissioning ba77170e03 10 2017 15 3 3 4 Setting the date and time 3 See section 4 5 5...
Страница 117: ......