4.4.3 Displaying measurement data
You can view the measurement data during a measurement (LED flashes blue)
and after a measurement (LED flashes green).
2.
Start the measurement with
<M>
.
If the signal LED is active, it will flash blue.
AutoTemp
is displayed.
3.
Keep the sample bottle (with the OxiTop
®
(-i/-IDS) measuring head
screwed on) in a place that is controlled to 20°C (e.g. thermostat cabi-
net/box) for the specified number of days.
When the measuring temperature has been reached (function
Auto-
Temp
:
after at least 1 hour, at most approx. 3 hours), the OxiTop
®
(-i/-
IDS) measuring head automatically starts measuring the oxygen con-
sumption.
4.
Stir the sample continuously during the specified measuring duration
(stirring platform).
During the specified measuring duration, the OxiTop
®
(-i/-IDS) measur-
ing head automatically saves a measured value once every 24 hours.
5.
Read out the saved measured values after the specified measuring
duration has expired.
If the signal LED is active, it will flash green.
1.
When the display is switched off:
Switch on the display with any key.
The current value (during measurement) or the value last determined
(after the end of the measurement) is displayed.
2.
Using
<
>
, scroll through the values stored last.
The measured value can be assigned to the day via the index
(BOD0,1,2 ...7).
3.
You can display the measured values as a graphic with
<
_
>
.