Oxi 3210
Operation
33
ba75795e07
10/2021
Note
The sponge in the air calibration vessel must be moist (not wet). Leave
the sensor in the air calibration vessel for a time long enough to adjust.
4.4.5
Displaying calibration records
Displaying the
calibration record
The calibration record of the last calibration is to be found under the
menu item,
Calibration
/
Calibration record
. To open it in the measured
value display, press the
<CAL__>
key.
2
Put the D.O. sensor into the air calibration vessel.
3
Start the calibration with
<CAL>
.
The last calibration data (relative slope) is displayed.
4
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
5
Wait for the end of the AutoRead measurement or accept the
calibration value with
<ENTER>
.
The calibration record is displayed.
6
Using
<F1>
/
[Continue]
or
<ENTER>
, switch to the measured
value display.
O
2
0.88
25.1
°C
OxiCal
15.03.2014
08:00
Содержание WTW Oxi 3210
Страница 1: ...OPERATING MANUAL ba75795e07 10 2021 Oxi 3210 DISSOLVED OXYGEN D O METER...
Страница 2: ...2 ba75795e07 10 2021 Oxi 3210 Copyright 2021 Xylem Analytics Germany GmbH Printed in Germany...
Страница 5: ...Oxi 3210 Contents 5 ba75795d07 10 2021...
Страница 10: ...Overview Oxi 3210 10 ba75795e07 10 2021...
Страница 14: ...Safety Oxi 3210 14 ba75795e07 10 2021...
Страница 41: ...Oxi 3210 Maintenance cleaning disposal 41 ba75795e07 10 2021...
Страница 43: ...Oxi 3210 What to do if 43 ba75795e07 10 2021...
Страница 46: ...Technical data Oxi 3210 46 ba75795e07 10 2021...
Страница 50: ...Lists Oxi 3210 50 ba75795d07 10 2021...
Страница 51: ......