18
ba15362d04
12/2021
Commissioning
NO 800
Commissioning
Combination elec-
trode NO 800
Conditioning, calibration, measurement
General informa-
tion
When operating the electrode ensure that
the filling opening (1) for the bridge electrolyte is open
the inner junctions (3) are covered with bridge electrolyte
no air bubbles are in the bridge electrolyte
the depth of immersion is within the optimum range:
Before measuring
1
Remove the storage vessel.
2
Fill ELY/BR/503/N bridge electrolyte into the filling opening so that
the inner junctions are covered with bridge electrolyte.
3
Rinse the combination electrode with deionized water.
4
Wipe the shaft using a clean paper towel.
Minimum depth of
immersion
The outer junction (4) must be covered
Maximum depth of
immersion
Approx. 1 cm below the fluid level (2) of the
bridge electrolyte
1
2
3
4
1
Condition every ion sensitive electrode in standard solution before
using it.
Содержание wtw NO 800 DIN
Страница 2: ...2 ba15362d04 12 2021 NO 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 5: ...BEDIENUNGSANLEITUNG ba15362d04 12 2021 NO 800 DIN NITRAT SELEKTIVE ELEKTRODE BEDIENUNGSANLEITUNG...
Страница 6: ...ba15362d04 12 2021 NO 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 14: ...14 ba15362d04 12 2021 Kalibrierkurve einer Nitrat selektiven Elektrode NO 800...
Страница 15: ...OPERATING MANUAL ba15362d04 12 2021 NO 800 DIN NITRATE SENSITIVE ELECTRODE OPERATING MANUAL...
Страница 16: ...ba15362d04 12 2021 NO 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 23: ...NO 800 What to do if ba15362d04 12 2021 23 Cable length Plug DIN plug 1 m...
Страница 24: ...24 ba15362d04 12 2021 What to do if NO 800...
Страница 25: ...MODE D EMPLOI ba15362f04 12 2021 NO 800 DIN LECTRODE S LECTIVE DE NITRATE MODE D EMPLOI...
Страница 26: ...ba15362f04 12 2021 NO 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 34: ...34 ba15362f04 12 2021 Courbe de calibration d une lectrode s lective de nitrate NO 800...
Страница 35: ...MANUAL DE INSTRUCCIONES ba15362s04 12 2021 NO 800 DIN ELECTRODO SELECTIVO PARA NITRATO MANUAL DE...
Страница 36: ...ba15362s04 12 2021 NO 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 44: ...44 ba15362s04 12 2021 Especificaciones t cnicas NO 800...
Страница 45: ......