NitraVis 70x IQ
Indexes
51
ba75951e03
05/2017
9.2
Status info
The status info is a coded piece of information on the current status of a sensor.
Each sensor sends this status info to the controller. The status info of sensors
consists of 32 bits, each of which can have the value 0 or 1.
Status info,
general structure
The bits 0 - 15 are reserved for general information.
The bits 16 - 21 are reserved for internal service information.
You obtain the status info:
via a manual query in the
Einstellungen/Settings
/
Service
/
List of all compo
-
nents
menu (see system operating manual)
by an automated query
– from a superordinate process control (e. g. when connected to the Profi
-
bus)
– of the IQ Data Server (see operating manual of the IQ S
ENSOR
N
ET
soft
-
ware pack)
Status-Info
NitraVis 70x IQ
0 1 2 3 4 5 6 7
8 9 10 11 12 13 14 15
1 0 0 0 0 0 0 0
0 0 0 0 0 0 0 0 (general)
0 0 0 0 0 0 0 0
0 0 0 0 0 0 0 0 (internal)
16 17 18 19 20 21 22 23
24 25 26 27 28 29 30 31
The evaluation of the status info, e.g. in the case of an automated
query, has to be made individually for each bit.
Status bit
Explanation
Bit 0
Component hardware defective
Bit 1
Component hardware defective
xxx
Bit 2
Optical measuring range exceeded
Bit 3-31
-
Содержание wtw NitraVis 705 IQ
Страница 1: ...OPERATING MANUAL ba75951e03 05 2017 NitraVis 701 IQ NitraVis 705 IQ OPTICAL SENSOR FOR NITRATE...
Страница 2: ...2 ba75951e03 05 2017 NitraVis 70x IQ Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Страница 54: ...Appendix Glossary NitraVis 70x IQ 54 ba75951e03 05 2017...
Страница 55: ......